R2 relaxometry based MR imaging for estimation of liver iron content: A comparison between two methods
Juan S Calle-Toro1, Christian A Barrera1, Dmitry Khrichenko1
1Division of Body Imaging, Department of Radiology, Children's Hospital of Philadelphia, 3401 Civic Center Blvd, Philadelphia, PA, 19104, USA.
This study shows R2 relaxometry MRI is a reproducible method for estimating liver iron concentration (LIC). Both in-house analysis and FDA-approved commercial software provide comparable results for LIC assessment.
Area of Science:
- Radiology
- Medical Imaging
- Biophysics
Background:
- Accurate assessment of liver iron concentration (LIC) is crucial for managing iron overload disorders.
- R2 relaxometry Magnetic Resonance Imaging (MRI) offers a non-invasive method for LIC estimation.
- Comparing in-house analysis with commercial, FDA-approved software is essential for clinical validation.
Purpose of the Study:
- To compare the reproducibility and accuracy of R2 relaxometry MRI for liver iron concentration (LIC) estimation.
- To evaluate the consistency between in-house analysis methods and FDA-approved third-party commercial results.
Main Methods:
- Utilized a 1.5T MRI scanner with multi-echo spin-echo sequences.
- Calculated R2 relaxivity values by analyzing regions of interest in the liver.
- Assessed the correlation and agreement between in-house R2-based LIC estimation and FerriScan® results using correlation coefficients and Bland-Altman plots.
Main Results:
- A total of 474 studies from 175 patients (age range 1-18 years) were analyzed.
- Excellent linearity was observed between proprietary and in-house R2 methods (r=0.87, p<0.001).
- Liver iron concentration (LIC) values reported by FerriScan® showed strong agreement with in-house R2-estimated LIC (r=1.0, p<0.001).
Conclusions:
- R2 relaxometry MRI is a reproducible technique for estimating liver iron concentration.
- Both proprietary FDA-approved commercial software and in-house analysis methods yield consistent results for LIC estimation.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Stereotype Content Model
What are Estimates?
The estimate for the mean of a sample is denoted by ͞x, whereas the mean of the population is designated as μ. Further, parameters such...
One-Compartment Open Model: Wagner-Nelson and Loo Riegelman Method for ka Estimation
On...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...