Comparison of Icare Rebound Tonometer and Perkins Applanation Tonometer in Community Eye Screening
Siew Leng Ting1, Lik Thai Lim1, Chor Yau Ooi2
1Department of Ophthalmology, Faculty of Medicine and Health Sciences, Universiti Malaysia Sarawak, Malaysia.
Purpose:
To compare the measurement of intraocular pressure (IOP) of Icare rebound tonometer and Perkins applanation tonometer (PAT) during community eye screening and to assess the agreement between these 2 instruments.
Design:
A cross-sectional, non-interventional study.
Methods:
The IOP measurements by handheld Icare rebound tonometer (Finland) were first performed by a primary care physician. Then the IOP was measured using Perkins Mk3 applanation tonometer (Haag-Streit, UK) by an ophthalmologist who was masked to previous readings from the Icare rebound tonometer. The mean IOP measured by each tonometer was compared. Pearson correlation coefficient was used to explore the correlation between the IOP measurements of the 2 instruments. The level of agreement between them was assessed using the Bland and Altman method.
Results:
A total of 420 left eyes were examined. The mean age of subjects was 38.6 ± 18.2 years. Approximately 67% of subjects were female. The mean IOP was 16.3 ± 4.0 mm Hg using Icare and 13.4 ± 2.3 mm Hg using PAT. Pearson correlation coefficient showed a moderate positive correlation between the 2 methods (r = +0.524, P < 0.001). Linear regression analysis revealed a slope of 0.28 with R² of 0.255. The mean difference between the 2 methods was 2.90 ± 3.5 mm Hg and the sample t-test revealed a statistically significant mean difference from 0 (P < 0.001). The 95% limits of agreement between the 2 methods were between -9.73 and 3.93 mm Hg.
Conclusions:
The handheld Icare rebound tonometer is a reasonably acceptable screening tool in community practices. However, Icare overestimated IOP with a mean of 2.90 mm Hg higher than the PAT. Thus, using Goldmann applanation tonometer as a confirmatory measurement tool of IOP is suggested.
Related Concept Videos
What are Populations and Communities?
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Community Based Intervention
Foundations of Community Mental Health Programs
Central to the success of community-based interventions is the...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...