Related Experiment Video
Updated: Jan 23, 2026

Amplification of Escherichia coli in a Continuous-Flow-PCR Microfluidic Chip and Its Detection with a Capillary Electrophoresis System
Published on: November 21, 2023
Amplification of lysozyme signal detected in capillary electrophoresis using mixed polymer brushes coating with
Mingxin Hou1, Miao Zhang1, Lijuan Chen2
1CAS Key Laboratory of Soft Matter Chemistry, Department of Polymer Science and Engineering, University of Science and Technology of China, Hefei, 230026, PR China.
Abstract:
In this work, a mixed polymer brushes based on poly(2-methyl-2-oxazoline) (PMOXA) and poly(acrylic acid) (PAA) coated capillary with switchable protein adsorption/desorption properties was developed and applied for on-line extraction and preconcentration of lysozyme. The study of electroosmotic flow (EOF) and fluorescence microscope showed that the inner surface charge of PMOXA/PAA mixed brush coated capillary displayed the switchable behavior toward the change of pH value and ionic strength (I), and PMOXA/PAA mixed brushes coated capillary could adsorb high amounts of lysozyme at pH 7 (I = 10-5 M), and the most of adsorbed lysozyme could then be desorbed at pH 3 (I = 10-1 M). Subsequently, this coated capillary with switchable lysozyme adsorption/desorption ability was applied for on-line extraction and preconcentration of lysozyme during capillary electrophoresis (CE) performance. Under the process of on-line preconcentration, the detection signal (peak area) of lysozyme obtained in PMOXA/PAA coated capillary was 26 times that obtained in bare capillary under normal CE while the contour chain length of PAA was 1.56 times that of PMOXA. Moreover, the value of low detection limit (LOD) of lysozyme using above coated capillary under on-line preconcentration method reached to 4.5 × 10-9 mg/mL, and 1 × 105-fold sensitivity enhancement was realized for lysozyme as compared with the bare capillary under normal CE.
More Related Videos
13:57Preparation and Friction Force Microscopy Measurements of Immiscible, Opposing Polymer Brushes
Published on: December 24, 2014
07:54Analysis of AtHIRD11 Intrinsic Disorder and Binding Towards Metal Ions by Capillary Gel Electrophoresis and Affinity Capillary Electrophoresis
Published on: August 22, 2018
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Polymers
Polymers
Surface Tension, Capillary Action, and Viscosity
The various IMFs between identical molecules of a substance are examples of cohesive forces. The molecules within a liquid are surrounded by other molecules and are attracted equally in all directions by the cohesive forces within the liquid. However, the molecules on the surface of a liquid are attracted only by about one-half as many molecules. Because of the unbalanced molecular attractions on the surface molecules, liquids contract to form a shape that minimizes the number...
Capillary Beds
Capillaries connect arterioles, small branches of arteries, to venules,...