Can venous cannula design influence venous return and negative pressure with a minimally invasive extracorporeal
Mark J Bennett1, Sian Hodgkiss2, Clinton T Lloyd3
1Cardiothoracic Anaesthesia, University Hospitals Plymouth NHS Trust, Plymouth, UK.
Introduction:
Recent advances to make cardiopulmonary bypass more physiological include the use of kinetic-assisted venous drainage but without a venous reservoir. Despite manipulation of intravascular volume and patient positioning, arterial flow is frequently reduced. Negative venous line pressures can be generated, which may elicit gaseous microemboli. We investigated the influence of venous cannula design on venous return and negative venous line pressures.
Methods:
In a single-centre, single-surgeon, prospective, randomized, double-blind trial, 48 patients undergoing isolated coronary artery, aortic valve or combined coronary artery and aortic valve surgery, with a minimally invasive circuit, were randomized to a conventional two-stage (2S) or three-stage venous cannula (3S), or to a three-stage venous cannula with additional 'fenestrated' ridges (F3S). Blood flow, venous line pressures and gaseous microemboli number and size were measured.
Results:
The pump flow achieved was the same between groups, but in each case fell below the target range of 2.2-2.4 L min-1 m-2. The three-stage cannula recorded significantly lower negative pressure than the other cannulae. The total count and volume of gaseous emboli detected with the F3S cannulae was very high in some cases, with wide heterogeneity.
Discussion:
The low negative pressures recorded with three-stage cannula, despite having a larger drainage orifice area, suggest that negative pressure may be more influenced by lumen diameter and vena cava collapse rather than drainage hole size. The additional fenestrations resulted in flow characteristics and negative pressures similar to the larger two-stage cannula but are associated with generation of gaseous microemboli.
Related Concept Videos
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system...
Venous Thrombosis I: Introduction
Venous Thrombosis III: Interprofessional Care
Venous Thrombosis IV: Nursing Management
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Biopharmaceutical Factors Influencing Drug Product Design: Overview