Comparison between the new fully automated viscoelastic coagulation analysers TEG 6s and ROTEM Sigma in trauma
Bernhard Ziegler1, Wolfgang Voelckel, Johannes Zipperle
1From the Departement of Anesthesiology and Intensive Care Medicine, University Hospital of Salzburg (BZ), Departement of Anesthesiology and Intensive Care Medicine, AUVA Trauma Centre Salzburg, Academic Teaching Hospital of the Paracelsus Medical University, Salzburg (WV, HS), Ludwig Boltzmann Institute for Experimental and Clinical Traumatology, AUVA Trauma Research Centre, Vienna, Austria (JZ, HS) and Department of Anesthesiology, RWTH Aachen University Hospital, Aachen, Germany (OG).
Background:
Viscoelastic coagulation testing is increasingly used to diagnose trauma-induced coagulopathy. Two fully automated analysers, TEG 6s and ROTEM Sigma, were launched recently. No previous studies have compared these devices in trauma paients.
Objective:
The aim of this study was to evaluate whether both fully automatic devices deliver comparable results.
Design:
Prospective observational study.
Setting:
Level one trauma centre from August 2017 to September 2018.
Patients:
A total of 105 blood samples from 67 trauma patients were analysed simultaneously on TEG 6s and ROTEM Sigma.
Main Outcome Measures:
TEG 6s assays kaolin (CK), RapidTEG (CRT), kaolin with heparinase (CKH) and functional fibrinogen were compared with ROTEM Sigma assays INTEM, EXTEM, HEPTEM and FIBTEM. TEG 6s functional fibrinogen level was compared with plasma fibrinogen concentration, measured using the Clauss method. Correlations were classified as weak (Spearman correlation coefficient 0.20 to 0.39), moderate (0.40 to 0.59), strong (0.60 to 0.79) or very strong (≥0.80).
Results:
The TEG 6s parameters reaction time, kinetic time and α-angle (CK, CRT and CKH assays) mostly showed strong correlations with the corresponding ROTEM parameters clotting time, clot formation time and α-angle (INTEM, EXTEM and HEPTEM assays). The exceptions were CRT reaction time vs. EXTEM clotting time, and CK α-angle vs. INTEM α-angle, which correlated moderately. Absolute values for many of these parameters showed significant differences between the two devices. Very strong correlations and similar absolute values were observed between TEG 6s maximum amplitude (CRT, CK and CKH assays) and ROTEM maximum clot firmness (EXTEM, INTEM and HEPTEM assays). Correlations were also very strong for functional fibrinogen maximum amplitude vs. FIBTEM maximum clot firmness and functional fibrinogen level vs. Clauss fibrinogen concentration, but absolute values were significantly different.
Conclusion:
Strong to very strong correlations were observed between corresponding TEG 6s and ROTEM Sigma parameters. However, absolute values showed significant differences for most of the measurements.
More Related Videos
Related Concept Videos
Observational Studies
There are three types of observational studies – Prospective, retrospective, and cross-sectional.
Prospective Study
Prospective studies, also known as longitudinal or cohort studies, are carried out by collecting future data from groups sharing similar characteristics. One...
Coagulation
Coagulation
During the coagulation phase, clotting factors, or procoagulants, play a vital role in initiating and progressing the coagulation cascade. This cascade is a series of reactions...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Naturalistic Observations
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...