Related Experiment Video
Updated: Jan 23, 2026

In Silico Clinical Trials for Cardiovascular Disease
Published on: May 27, 2022
Impact of electrical pulse cut mode during endoscopic papillectomy: Pilot randomized clinical trial
Eisuke Iwasaki1, Kazuhiro Minami1, Takao Itoi2
1Division of Gastroenterology and Hepatology, Department of Internal Medicine, Keio University School of Medicine, Tokyo, Japan.
Objective:
Endoscopic papillectomy is increasingly being used for ampullary adenoma treatment. However, it remains challenging despite increased safety with treatment advances. The ideal power output and electrosurgical current mode for mucosal resection are not established. We aimed to identify the ideal electrical pulse for use during resection.
Methods:
This pilot randomized, single-blind, prospective, multicenter trial, recruited patients with ampullary adenomas and conventional anatomy who were scheduled to undergo endoscopic papillectomy. Endoscopic treatment was performed using a standardized algorithm and patients were randomized for endoscopic papillectomy with Endocut or Autocut. The primary outcome was the incidence of delayed bleeding. Incidence of procedure-related pancreatitis, successful complete resection, pathological findings, and other adverse events were secondary endpoints.
Results:
Sixty patients were enrolled over a 2-year period. The incidences of delayed bleeding (13.3% vs. 16.7%, P = 1.00) and pancreatitis (27% vs. 30%, P = 0.77) were similar between both groups. The rate of crush artifacts was higher in the Endocut than in the Autocut group (27% vs. 3.3%, P = 0.03). Immediate bleeding when resecting tumors greater than 14 mm in diameter was more common in the Autocut than in the Endocut group (88% vs. 46%, P = 0.04).
Conclusions:
The Autocut and Endocut modes have similar efficacy and safety for endoscopic papillectomy. The Endocut mode may prevent immediate bleeding in cases with large tumor sizes, although it causes more frequent crush artifacts.
Registry And The Registration Number:
The Japanese UMIN Clinical Trials Registry (UMIN-CTR: 000021382).
Related Concept Videos
Clinical Trials
There are four phases in a clinical trial. A phase one...
Clinical Trials: Overview
Statistical Software for Data Analysis and Clinical Trials
Pilot and Numeric Relaying
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Pulse
The pulse serves as a clinical...