Technical Note: Alternating clinical usage of the integral quality monitor transmission detector
Götz Kohler1, Nicolas Hanauer1
1Clinic of Radiotherapy and Radiation Oncology, University Hospital Basel, 4031, Basel, Switzerland.
Purpose:
The purpose of the study was the implementation of a method to use the integral quality monitor (IQM) transmission detector for occasional, alternating usage during patient treatment with intensity modulated radiotherapy. Due to attenuation, the transmission detector must be taken into account during the planning process. The proposed workflow is based on determining a dynamic transmission factor (dTF) required to scale the total number of MU of the original radiotherapy (RT) plan. Thus a very similar radiation therapy plan is obtained that can be used with the IQM detector.
Methods:
Ten clinically applied volumetric modulated arc therapy plans were delivered at two beam qualities. A dTF is calculated from each RT plan for which a collapsed RT plan was created for verification using a two-dimensional array with and without the IQM detector. The total number of MU of the original RT plan was scaled by the inverse of the dTF to obtain the modified RT plan for clinical use with the IQM detector. Validation was performed with an electronic three-dimensional phantom and via gamma analysis using strict criteria of 1%/1 mm.
Results:
Except for one outlier, the gamma pass rate between the original RT plan without IQM and the modified RT plan with IQM was always above 99.5%. The variations of the dTF were smaller than 1% for all tested RT plans.
Conclusions:
The results show that the proposed workflow can be used clinically. Thus the IQM transmission detector can also be used occasionally for online verification of RT plans.
More Related Videos
05:54Stereo-Electro-Encephalo-Graphy SEEG With Robotic Assistance in the Presurgical Evaluation of Medical Refractory Epilepsy: A Technical Note
Published on: June 13, 2016
05:58Application of an Amplitude-integrated EEG Monitor Cerebral Function Monitor to Neonates
Published on: September 6, 2017
Related Concept Videos
Alternative RNA Splicing
There are five types of alternative RNA splicing that vary in the ways the pre-mRNA segments are removed or retained in the mature mRNA. The first...
Alternative RNA Splicing
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Null and Alternative Hypotheses
The null hypothesis, denoted by H0 is a statement of no difference between the variables—they are not related. This can often be considered the status quo. As a result if you cannot accept the null, it requires some action.
The alternative hypothesis, denoted by H1 or Ha, is a claim about the...