Related Experiment Video
Updated: Jan 23, 2026

Simulation of a Scaled Assembly Process with Collaboration of a Robotic Arm and Monitoring through a Vision System for Quality Control
Published on: August 29, 2025
An improved treatment planning and quality assurance process for Collaborative Ocular Melanoma Study eye plaque
R A Weersink1, S Patterson2, H Ballantyne2
1Department of Radiation Oncology, University of Toronto, Toronto, Canada; Radiation Medicine Program, Princess Margaret Cancer Centre, Toronto, Canada; Institute of Biomaterials and Biomedical Engineering, University of Toronto, Toronto, Canada; Techna Institute, University Health Network, Toronto, Canada.
Purpose:
To develop a treatment planning platform for episcleral Collaborative Ocular Melanoma Study plaque therapy in an established treatment planning software and to improve an existing quality assurance (QA) process for nonuniformly loaded plaques that measures air kerma strengths (AKSs) and loading profile.
Materials And Methods:
Treatment planning is performed in Pinnacle using scripts that let the planner choose plaque size and notching. Scripts load seed positions for each plaque and five source groups corresponding to available stock seeds that can be placed into each seed position. Contours are loaded that display the model eye and the plaque itself. Plaque QA is performed using a modification of our previous pinhole apparatus by replacing x-ray film exposure with an optical camera and scintillating film system. The captured image is processed to remove background and to correct the intensity of seeds on the plaque periphery. Measured total optical counts provide an estimate of total plaque AKS.
Results:
Treatment planning of eye plaques using Pinnacle, in conjunction with our stock inventory of seeds, is established as standard practice at our center. Planned plaques can vary from uniformly loaded to asymmetrically loaded notched plaques. Using the optical camera system for assessment of the seed loadings has decreased QA time from 40 min/plaque to 10 min/plaque. Total AKS of each plaque can be measured using the optical camera with an accuracy of 10%.
Conclusions:
Treatment planning is performed on a Health Canada-approved software that accommodates nonuniform plaque loading. Optical imaging of the plaque provides absolute total AKS and the relative seed arrangement in the plaque.
Related Concept Videos
Quality Assurance
Role of Communication in the Nursing Process II: Planning and Implementation
Nursing Process for Patient and Caregiver Teaching II: Planning and Implementation
Chronic Pancreatitis II: Collaborative Care
Assessment:
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Planning Nursing Care I