Use of Rabbit Eyes in Pharmacokinetic Studies of Intraocular Drugs
Published on: July 23, 2016
Changes in intraocular pressure after intraocular eye surgery-the influence of measuring technique
Hrvoje Kovacic1, Roger C W Wolfs1, Emine Kılıç1
1Department of Ophthalmology, Erasmus Medical Center, Rotterdam 3000 CA, the Netherlands.
Aim:
To investigate the changes in intraocular pressure (IOP) before and after intraocular surgery measured with Goldmann applanation tonometry (GAT) and pascal dynamic contour tonometry (PDCT), and assessed their agreement.
Methods:
Patients who underwent trans pars plana vitrectomy (TPPV) with or without cataract extraction (CE) were included. The IOP was measured in both eyes with GAT and PDCT pre- and postoperatively, where the non-operated eyes functioned as control.
Results:
Preoperatively, mean IOP measurements were 16.3±6.0 mm Hg for GAT and 12.0±2.8 mm Hg for PDCT for the operated eyes. Postoperatively, the mean IOP dropped to 14.3±5.6 mm Hg for GAT (P=0.011) and rose up to 12.7±2.6 mm Hg for PDCT (P=0.257). Bland-Altman analysis showed a poor agreement between GAT and PDCT with a mean difference of 2.9 mm Hg preoperatively and 95% limits of agreement ranging from -3.2 to 9.0 mm Hg. Postoperatively, the mean difference was 1.2 mm Hg with 95% limits of agreement ranging from -8.3 to 10.7 mm Hg. There were no significant differences between the TPPV and TPPV+CE group, except when measured with PDCT postoperatively (P=0.012).
Conclusion:
The IOP is reduced after surgery when measured with GAT and remained stable when measured with PDCT. However, the agreement between GAT and PDCT is poor. Although PDCT may be a more accurate predictor of the true IOP, it seems less suitable for daily use in the clinical practice.
Related Concept Videos
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Vapor Pressure