Related Experiment Video
Updated: Jan 22, 2026

Highly Sensitive and Rapid Fluorescence Detection with a Portable FRET Analyzer
Published on: October 1, 2016
High-throughput fluorescence polarization immunoassay by using a portable fluorescence polarization imaging analyzer
Osamu Wakao1, Ken Satou2, Ayano Nakamura1
1Graduate School of Chemical Sciences and Engineering, Hokkaido University, Kita 13 Nishi 8, Kita-ku, Sapporo 060-8628, Japan.
A portable, low-cost fluorescence polarization analyzer enables high-throughput mycotoxin detection. This system facilitates simultaneous testing of 96 samples or over 500 samples, paving the way for on-site diagnostics.
Area of Science:
- Analytical Chemistry
- Biotechnology
- Biophysics
Background:
- Mycotoxin detection is crucial for food safety and requires sensitive, high-throughput methods.
- Existing methods can be costly and time-consuming, limiting on-site applications.
- Fluorescence polarization immunoassays (FPIAs) offer a sensitive detection principle.
Purpose of the Study:
- To develop and demonstrate a portable, cost-effective fluorescence polarization analyzer for high-throughput mycotoxin detection.
- To validate the system's capability for simultaneous and high-throughput FPIA measurements.
- To assess the potential for automated sample reaction and detection.
Main Methods:
- High-throughput FPIAs were performed using a portable fluorescence polarization (FP) analyzer integrated with a microdevice.
- Simultaneous FPIA measurements were conducted for 8 deoxynivalenol (DON) concentrations across 12 chambers (96 samples).
- High-throughput FPIA measurements were performed for single DON concentrations in over 500 chambers using FP imaging.
Main Results:
- The portable FP analyzer, costing under $5000 and with a 1 nL sample volume, successfully enabled simultaneous FPIA for 96 samples.
- High-throughput FPIA measurements for over 500 samples were demonstrated using FP imaging.
- The system's portability and low cost are suitable for on-site measurement and point-of-care testing.
Conclusions:
- The developed portable FP analyzer offers a cost-effective, high-throughput solution for mycotoxin detection.
- FP imaging enables simultaneous and high-throughput FPIA, expanding applications in diagnostics and molecular interaction studies.
- Future automation of sample reaction and detection is anticipated, further enhancing its utility.
More Related Videos
11:17Monitoring Changes in Membrane Polarity, Membrane Integrity, and Intracellular Ion Concentrations in Streptococcus pneumoniae Using Fluorescent Dyes
Published on: February 17, 2014
09:38Analyzing the Function of Small GTPases by Microinjection of Plasmids into Polarized Epithelial Cells
Published on: May 31, 2011
Related Concept Videos
Group Polarization
Molecular Shape and Polarity
Polarity of the Cytoskeleton
Bond Polarity, Dipole Moment, and Percent Ionic Character
Polar Covalent Bonds
Dielectric Polarization in a Capacitor