Iterative CT reconstruction in abdominal low-dose CT used for hybrid SPECT-CT applications: effect on image quality,
Oliver S Grosser1, Juri Ruf2, Dennis Kupitz1
1Department of Radiology and Nuclear Medicine, University Hospital Magdeburg, Magdeburg, Germany.
Background:
Iterative computed tomography (CT) image reconstruction shows high potential for the preservation of image quality in diagnostic CT while reducing patients' exposure; it has become available for low-dose CT (LD-CT) in high-end hybrid imaging systems (e.g. single-photon emission computed tomography [SPECT]-CT).
Purpose:
To examine the effect of an iterative CT reconstruction algorithm on image quality, image noise, detectability, and the reader's confidence for LD-CT data by a subjective assessment.
Material And Methods:
The LD-CT data were validated for 40 patients examined by an abdominal hybrid SPECT-CT (U = 120 kV, I = 40 mA, pitch = 1.375). LD-CT was reconstructed using either filtered back projection (FBP) or an iterative image reconstruction algorithm (Adaptive Statistical Iterative Reconstruction [ASIR]®) with different parameters (ASIR levels 50% and 100%). The data were validated by two independent blinded readers using a scoring system for image quality, image noise, detectability, and reader confidence, for a predefined set of 16 anatomic substructures.
Results:
The image quality was significantly improved by iterative reconstruction of the LD-CT data compared with FBP (P ≤ 0.0001). While detectability increased in only 2/16 structures (P ≤ 0.03), the reader's confidence increased significantly due to iterative reconstruction (P ≤ 0.002). Meanwhile, at the ASIR level of 100%, the detectability in bone structure was highly reduced (P = 0.003).
Conclusion:
An ASIR level of 50% represents a good compromise in abdominal LD-CT image reconstruction. The specific ASIR level improved image quality (reduced image noise) and reader confidence, while preserving detectability of bone structure.
More Related Videos
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography