Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
Feasibility of An Entirely Extracardiac, Minimally Invasive,Temporary Pacing System
Anne-Floor B E Quast1, Niek E G Beurskens1, Adrian Ebner2
1Amsterdam UMC, University of Amsterdam, Heart Center, Department of Clinical and Experimental Cardiology, Amsterdam Cardiovascular Sciences, the Netherlands (A.-F.B.E.Q., N.E.G.B., J.T.V., R.E.K., M.C.B.).
Background:
A completely extracardiac pacing system provides the potential for clinical advantages over existing device alternatives that require intravascular, endocardial, or epicardial contact. Preliminary studies evaluating the feasibility of cardiac pacing with a lead in the anterior mediastinum, outside the pericardium and circulatory system have been completed. These studies examined (1) the anatomic access route, (2) the usability of a delivery tool to facilitate lead placement, and (3) the pacing performance of the extracardiac lead.
Methods:
Feasibility evaluations included (1) a retrospective computed tomography analysis to characterize anatomic variations related to lead access, (2) accessing the anterior mediastinum in cadavers and human subjects using a custom delivery tool, and (3) acute clinical pacing performance.
Results:
Major findings: (1) A total of 166 (95%) out of 174 patients had a viable lead access path through the fourth, fifth, or sixth intercostal space. (2) Access to the targeted implant location using a delivery tool was successful in all 5 cadavers and 3 humans without use of fluoroscopy and with an average lead delivery time of 121±52 s. No damage to the lung, pericardium, heart, or internal thoracic vessels occurred. (3) Pacing performance was tested in 6 human subjects showing a threshold voltage of 4.7 V (2.7-6.7), threshold pulse width of 1.8 ms (1.0-2.5), and an impedance of 1205 Ω (894-1786). R-wave amplitudes measured 9.6 mV (5.6-12.0).
Conclusions:
Results support the feasibility for this completely extracardiac pacing method in a heterogeneous patient population, using a minimally invasive, parasternal, delivery approach and with adequate sensing and thresholds suited for temporary pacing.
More Related Videos
Related Concept Videos
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Colors and Magnetism
When atoms or molecules absorb light at the proper frequency, their electrons are excited to higher-energy orbitals. For many main group atoms and molecules, the absorbed photons are in the ultraviolet range of the electromagnetic spectrum, which cannot be detected by the human eye. For coordination compounds, the energy difference between the d orbitals often allows photons in the visible range to be absorbed and emitted, which is seen as colors by the human...
Intermolecular Forces
Van der Waals Interactions
Hormonal Regulation