Diffusion MRI: Assessment of the Impact of Acquisition and Preprocessing Methods Using the BrainVISA-Diffuse Toolbox
Lucile Brun1, Alexandre Pron1, Julien Sein1
1Institut de Neurosciences de La Timone, Aix-Marseille University, CNRS, UMR 7289, Marseille, France.
Acquiring a b0 volume with reversed phase-encoding (RPE) offers the best balance between scan time and diffusion MRI data quality. This method improves geometric accuracy and tensor estimation without image smoothing, outperforming other preprocessing pipelines.
Area of Science:
- Neuroimaging
- Diffusion MRI analysis
- Medical physics
Background:
- Diffusion MRI is susceptible to geometric distortions from motion, eddy currents, and magnetic susceptibility variations.
- Existing correction methods vary in acquisition settings and data quality outcomes.
- The trade-off between scan time and preprocessing data quality remains unevaluated.
Purpose of the Study:
- To investigate the impact of six preprocessing pipelines on diffusion MRI analysis outcomes.
- To evaluate the ratio between scan time and preprocessed data quality.
- To develop a toolbox (Diffuse) guiding users to optimal preprocessing pipelines.
Main Methods:
- Comparison of six preprocessing pipelines using diffusion MR images from 20 HCP dataset subjects.
- Evaluation criteria included geometric accuracy, tensor model estimation, white matter indices, and pathway dispersion.
- A pipeline with reversed phase-encoding (RPE) acquisition served as the quality reference.
Main Results:
- Eddy-current correction enhances tensor fitting, particularly in the corpus callosum.
- Field maps alone for susceptibility correction require smoothing, artificially reducing tensor-fitting error.
- Acquiring a b0 volume with RPE at equivalent scan times yields superior geometry reconstruction and local tensor quality without smoothing.
Conclusions:
- Diffusion MRI metrics are highly dependent on acquisition settings and preprocessing methods.
- Acquiring b0 with RPE presents the optimal scan time/data quality compromise for diffusion MRI.
- The Diffuse toolbox aids in selecting appropriate preprocessing strategies based on input data.
Related Concept Videos
Diffusion
Diffusion
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Facilitated Diffusion
In this process, substrates such as organic compounds and ions interact with a transporter on one side, triggering conformational changes in proteins that enable...
Behavior of Gas Molecules: Molecular Diffusion, Mean Free Path, and Effusion
Protein Diffusion in the Membrane