Transcatheter aortic valve deployment influences neo-sinus thrombosis risk: An in vitro flow study
Immanuel David Madukauwa-David1, Vahid Sadri1, Norihiko Kamioka2
1Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology, Atlanta, Georgia.
Objectives:
We investigated the impact of (transcatheter heart valve) THV expansion at the level of the native annulus and implant depth on valve performance and neo-sinus flow stasis.
Background:
Flow stasis in the neo-sinus is one of the identified risk factors of THV thrombosis.
Methods:
A 29 mm CoreValve and 26 mm SAPIEN 3 were deployed under different expansions (CoreValve, SAPIEN 3) and implant depths (CoreValve) within a patient-derived aortic root in a pulse duplicator. Fluorescent dye was injected during diastole into the neo-sinus and imaged over 20 cardiac cycles. Washout times were computed as a measure of flow stasis for each deployment.
Results:
The 10% CoreValve under-expansion improved neo-sinus washout over full expansion by 8% (p < .001), and higher CoreValve implant depth improved neo-sinus washout (p < .001). The 10% SAPIEN 3 under-expansion improved neo-sinus washout by 23% (p < .001). Under-expansion of both valve types caused higher pressure gradients and smaller effective orifice areas than full expansion.
Conclusions:
Neo-sinus flow stasis is influenced by THV expansion and implant depth (CoreValve). The 10% valve under-deployment (oversizing) may facilitate reduced flow stasis in the neo-sinus with minimal increase in pressure gradients. This strategy may be helpful for patient anatomies, which are in-between transcatheter valve sizes.
More Related Videos
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Venous Thrombosis I: Introduction
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Relative Risk
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism,...
Factors Affecting the Risk of Infection
The integrity and count of the white blood cells help the body resist pathogens and fight infection. When impaired, it reduces the body's resistance to pathogens. The acidic pH levels of the gastrointestinal, genitourinary tracts, and skin...