[Comparison of biometry with the Pentacam AXL, IOLMaster 700 and IOLMaster 500 in cataract patients]
1Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Lab., Beijing 100730, China.
Abstract:
Objective: To compare biometry with new biometers of Pentacam AXL and IOLMaster 700 and the widely used biometer of IOLMaster 500 in cataract patients. Methods: Cross-sectional study. A total of 223 eyes of 147 cataract patients from Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University during 4-8 December 2017 were recruited. There were 67 males and 80 females with an age of (64±11) years.The axial length (AL), mean keratometry (Km), anterior chamber depth (ACD) and white-to-white corneal diameter (WTW) from each device were recorded. The difference and agreement between the measurements were evaluated by ANOVA, LSD test, intraclass correlation coefficient and Bland-Altman analysis. Results: The success rate of AL measurement was 85.2% (190/223) with the Pentacam AXL, 96.9% (216/223) with the IOLMaster 700 and 90.6% (202/223) with the IOLMaster 500. There was no difference among devices in measuring the AL, Km and ACD (all P>0.05). Significant difference was found in WTW [(11.36±0.42) vs. (11.69±0.45) vs. (11.45±0.42) mm; F=34.696, P=0.000]. Intraclass correlation coefficient was 0.859-1.000, and Bland-Altman analysis showed good agreement among three devices. Conclusions: The new biometers of Pentacam AXL and IOLMaster 700 and the widely used IOLMaster 500 show no difference in measuring AL, Km and ACD. All three devices show good agreement in biometry of cataract patients. (Chin J Ophthalmol, 2019, 55: 515-521).
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
07:53Establishment of a Clinically Relevant Ex Vivo Mock Cataract Surgery Model for Investigating Epithelial Wound Repair in a Native Microenvironment
Published on: June 5, 2015
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits
Molecular Comparison of Gases, Liquids, and Solids