Smartphone Application with Virtual Reality Goggles for the Reliable and Valid Measurement of Active Craniocervical
Ke-Vin Chang1,2,3, Wei-Ting Wu1,2, Mei-Chu Chen1
1Department of Physical Medicine and Rehabilitation, National Taiwan University Hospital, Bei-Hu Branch 10845, Taiwan.
Objective:
This study aimed to determine the intra-rater and inter-rater reliability and validity of a hybrid device, combining virtual reality goggles, a magnetometer and an inclinometer application for smartphones, to measure craniocervical range.
Summary Of Background Data:
Accurate evaluation of craniocervical range of motion is important for early detection of certain diseased conditions and monitoring the progress of interventions. The universal goniometer is widely used for the measurement but it requires experienced practitioners. Whether a combination of virtual reality goggles and smartphone applications can provide the same or better performance compared with the goniometer is still unknown.
Methods:
Forty-one healthy adults from the department of physical medicine and rehabilitation were recruited for craniocervical range examination (flexion, extension, side-bending to the right or left and rotating to the right or left) by using the hybrid device and universal goniometer. Using the hybrid device, repeated measurements were performed twice by a primary rater and once by a second rater. The primary rater also conducted a measurement using the universal goniometer in the same cohort. The intra-rater and inter-rater reliability (intra-class correlation coefficient (ICC)) were calculated using the two-way random effect model, whereas the validity was examined by the Pearson correlation coefficient and Bland-and-Altman plot. The interval between the first and second sessions of the measurement for intra-rater reliability was set at 30 min.
Results:
Excellent intra-rater (ICC ≥ 0.925) and inter-rater (ICC ≥ 0.880) reliability was noted for the hybrid device. The minimal detectable changes from intra-observer and inter-observer comparisons ranged between 4.12° and 7.42° in all six directions. The Bland-and-Altman plot revealed small mean differences (≤1.68°) between the hybrid device and universal goniometer. Both instruments had highly correlated measurements of craniocervical motion (r values ≥ 0.918).
Conclusion:
For healthy participants, excellent intra-rater and inter-rater reliability was noted for the hybrid device, and the measurements were consistent with the universal goniometer measurements. Future studies are needed to examine whether the device can perform similarly for patients with neck disorders.
Related Concept Videos
Reliability and Validity
Dynamics Of Circular Motion: Applications
Application of Antiderivatives: Linear Motion
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
Variation: Normal Distribution, Range, and Standard Deviation
Virtual Work
In static equilibrium, a body can experience an imaginary or virtual movement, such as displacement or rotation. The virtual work done by a force is equal to the dot product of force and virtual displacement in the direction of the force. When it comes to virtually rotating a...