Urine protein detection by dipstick: No interference from alkalinity or specific gravity.
Jason L Robinson1, Allison A Venner1, Isolde Seiden-Long2
1Alberta Public Laboratories, Diagnostic and Scientific Research Centre, #9 3535 Research Way NW, Calgary, AB T2L 2K8, Canada; Department of Pathology and Laboratory Medicine, Cumming School of Medicine, Health Sciences Centre, Foothills Campus, University of Calgary, 3030 Hospital Drive NW, Calgary, AB T2N 4N1, Canada.
Tetrabromphenol blue dipsticks reliably detect proteinuria in patient urine within normal pH and specific gravity ranges. Alkalinity and high ionic strength can cause false results, but these are uncommon in physiological conditions.
Area of Science:
- Clinical Chemistry
- Urinalysis
- Diagnostic Accuracy
Background:
- Tetrabromphenol blue dye is used in urinalysis dipsticks for proteinuria detection.
- Manufacturers claim pH and specific gravity can cause false positive or negative results, but evidence is equivocal.
Purpose of the Study:
- To investigate the impact of urine pH and specific gravity on protein detection using tetrabromphenol blue dipsticks.
- To determine if common dipstick interferences are clinically relevant.
Main Methods:
- Three tetrabromphenol blue dipsticks (iChem 10SG, iChem Velocity, Multistix 8SG) were tested.
- Urine pools were adjusted for pH using NaOH, Na2CO3, or NaHCO3, and for specific gravity using NaCl.
- Protein, pH, specific gravity, electrolytes, and total CO2 were measured.
Main Results:
- Protein detection was unaffected by NaOH-induced alkalinization up to pH 10.9.
- False positive protein results occurred with Na2CO3 at pH 9.9 (>97 mmol/L total CO2) and NaHCO3 at pH 7.6 (>137 mmol/L total CO2).
- False negative protein results occurred when ionic strength exceeded 1100 mmol/L due to NaCl.
Conclusions:
- Tetrabromphenol blue dipsticks demonstrate robust proteinuria detection within physiological urine pH, total CO2, and ionic strength.
- The tested dipsticks are reliable for clinical use, as extreme conditions causing false results are rare in fresh patient samples.
Related Concept Videos
Specific Gravity of Aggregate
Bulk or gross specific gravity is calculated by taking the ratio of the mass of aggregates in the saturated surface-dry state to the total volume that includes both the solids and the voids within the...
Density, Specific Weight, Specific Gravity and Compressibility of Fluid
Specific weight represents the weight per unit volume and is calculated by multiplying...
Filtration and Urine Formation
RNA Interference
This process occurs naturally in cells, often through the activity of genomically-encoded microRNAs. Researchers can take advantage of this mechanism by introducing synthetic RNAs to deactivate specific genes for research or therapeutic purposes. For example, RNAi could be used...
Interference and Diffraction
Center of Gravity