Effect of microcollection tube fill volume on common acute care tests
Fangze Cai1, Isolde Seiden-Long1, Allison A Venner1
1Alberta Precision Laboratories, Diagnostic and Scientific Research Centre, #9 3535 Research Way NW, Calgary, AB T2L 2K8, Canada; Department of Pathology and Laboratory Medicine, Cumming School of Medicine, Health Sciences Centre, Foothills Campus, University of Calgary, 3030 Hospital Drive NW, Calgary, AB T2N 4N1, Canada.
Background:
Microcollection tubes are frequently used in pediatric phlebotomy. We performed a pilot study to determine what clinical biochemistry and hematology tests can be reported on different microcollection tube fill volumes.
Methods:
Blood was collected from 11 volunteers into Becton Dickinson (BD) Vacutainers® and microcollection tubes (BD Microtainers® and the Sarstedt Microvette® 300 FH) at different fill volumes (Filled: top line; Intermediate: second line; Short: third line. If there was no second or third line, 200 µL was used for short fills) for 36 clinical biochemistry tests and the complete blood count (CBC) with differential (23 components). At each fill volume, tests were strong candidates to report if they did not have statistically significant biases compared to results in Vacutainers®. Potential candidates had statistically significant biases that were small (median absolute bias < 25 % of total allowable error and less than desirable bias from biological variation). Tests were not candidates if biases were significant and large (median absolute bias ≥ 25 % of TEa or ≥ desirable bias). Biases that increased or decreased across concentration ranges invalidated reporting candidacy.
Results:
Twenty four clinical biochemistry tests were strong or potential candidates to report on all fill volumes, 7 were strong or potential candidates to report on some fill volumes and 5 were not candidates to report on any fill volumes. Seventeen CBC components were strong or potential candidates to report on all fill volumes, 2 were strong or potential candidates to report on some fill volumes and 4 were not candidates to report on any fill volumes.
Conclusions:
While most tests were valid to report on different fill volumes, some were not. We encourage laboratories to perform their own studies on fill volumes.
More Related Videos
Related Concept Videos
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Urodynamic Studies: Uroflowmetry
Sputum Studies II: Culture and Sensitivity
Sputum culture and sensitivity is a medical procedure used to diagnose bacterial infections in the respiratory tract and select the most appropriate antibiotics for treatment. This process involves analyzing sputum samples of thick and opaque secretions produced in the lungs and airways. These samples are collected from patients and then sent to the laboratory for analysis.
The test can identify various pathogens responsible for respiratory infections, including Streptococcus,...