Tissue response to applied loading using different designs of penile compression clamps
Joseph Mh Lemmens1, Jackie Broadbridge1, Margaret Macaulay1
1University of Southampton, School of Health Sciences, Southampton SO17 1BJ, UK.
Penile compression devices (PCDs) can impede blood flow and cause inflammation, but skin health recovers after removal. Applying a clamp for one hour with equal clamp-free time appears safe for managing urinary incontinence.
Area of Science:
- Urology
- Biomedical Engineering
- Dermatology
Background:
- Penile compression devices (PCDs) manage urinary incontinence (UI) post-prostatectomy.
- PCDs offer secure UI management but pose risks like pressure ulcers and impaired blood flow.
- Objective assessment of tissue health and safe PCD use guidelines are lacking.
Purpose of the Study:
- To evaluate physiological responses to existing PCDs.
- To assess the potential for pressure-induced injury from PCDs.
- To compare different PCD designs and inform safe usage recommendations.
Main Methods:
- Six men with post-prostatectomy UI participated in a controlled laboratory study.
- Four selected PCDs were tested in random order at effective pressures.
- Objective measures of skin injury, circulatory impedance, interface pressures, and inflammatory response (IL-1α) were recorded.
Main Results:
- PCD use demonstrated evidence of circulatory impedance under most test conditions.
- Interface pressures varied significantly between devices and participants (mean 137.4±69.7 mmHg).
- Elevated IL-1α levels indicated an inflammatory response after 10 minutes, returning to baseline 40 minutes post-removal.
Conclusions:
- All tested PCDs caused temporary tissue and blood flow compromise during 50-minute wear.
- While pro-inflammatory cytokines elevated, PCDs did not induce sustained irritation; skin health recovered post-removal.
- Applying a clamp for 1 hour with 1 hour free appears safe, but longer durations require further investigation.
Related Concept Videos
Behavior of Concrete Under Compressive Load
As the concrete specimen fractures under...
Design of Columns under a Centric Load
Euler's formula is applicable under the assumption that the column is a perfect, straight, homogenous prism, and it is operating...
Design of Columns under an Eccentric Load
Design Example: Frog Muscle Response
When the switch connecting the RL circuit is closed, a brief muscle contraction is observed. This is because, at a steady state, the inductor acts like a short...
Elastin is Responsible for Tissue Elasticity
Ligaments and tendons are made of dense regular connective tissue, but in ligaments not all fibers are parallel. Dense regular elastic tissue contains elastin fibers and...
Group Design