Fat suppression for ultrashort echo time imaging using a novel soft-hard composite radiofrequency pulse
Ya-Jun Ma1, Saeed Jerban1, Hyungseok Jang1
1Department of Radiology, University of California, San Diego, California.
Purpose:
To design a soft-hard composite pulse for fat suppression and water excitation in ultrashort echo time (UTE) imaging with minimal short T2 signal attenuation.
Methods:
The composite pulse contains a narrow bandwidth soft pulse centered on the fat peak with a small negative flip angle (-α) and a short rectangular pulse with a small positive flip angle (α). The fat magnetization experiences both tipping-down and -back with an identical flip angle and thus returns to the equilibrium state, leaving only the excited water magnetization. Bloch simulations, as well as knee, tibia, and ankle UTE imaging studies, were performed to investigate the effectiveness of fat suppression and corresponding water signal attenuation. A conventional fat saturation (FatSat) module was used for comparison. Signal suppression ratio (SSR), defined as the ratio of signal difference between non-fat-suppression and fat-suppression images over the non-fat-suppression signal, was introduced to evaluate the efficiency of the composite pulse.
Results:
Numerical simulations demonstrate that the soft-hard pulse has little saturation effect on short T2 water signals. Knee, tibia, and ankle UTE imaging results suggest that comparable fat suppression can be achieved with the soft-hard pulse and the FatSat module. However, much less water saturation is induced by the soft-hard pulse, especially for short T2 tissues, with SSRs reduced from 71.8 ± 6.9% to 5.8 ± 4.4% for meniscus, from 68.7 ± 5.5% to 7.7 ± 7.6% for bone, and from 62.9 ± 12.0% to 4.8 ± 3.2% for the Achilles tendon.
Conclusion:
The soft-hard composite pulse can suppress fat signals in UTE imaging with little signal attenuation on short T2 tissues.
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Toughness and Hardness of Aggregate
Classifying Matter by Composition
According to its composition, the matter can be classified into two broad categories — pure substances and mixtures.
A pure substance is a form of matter that has a constant composition throughout with uniform properties. For example, any sample of sucrose has the same composition and same physical properties, such as melting point, color, and sweetness, regardless of the source from which it is isolated.
A mixture is composed of two or...