High-throughput method to analyze tegafur and 5-fluorouracil in human tears and plasma using hydrophilic interaction
Ritsuko Shiokawa1,2, Xiao-Pen Lee2, Miho Yamada1
1Department of Ophthalmology, Showa University School of Medicine, 1-5-8 Hatanodai, Shinagawa-ku, Tokyo, 142-8555, Japan.
Rationale:
We developed a new high-throughput method to analyze tegafur (FT) and 5-fluorouracil (5-FU) in tear and plasma samples using hydrophilic interaction liquid chromatography (HILIC)/tandem mass spectrometry (MS/MS).
Methods:
The tear samples (10 μL) spiked with FT, 5-FU, and 5-chlorouracil (internal standard) were diluted using 40 μL of 2 M ammonium acetate and 250 μL of acetonitrile with 2% formic acid; 20 μL of plasma spiked with the two drugs and internal standard was diluted with 80 μL of 2 M ammonium acetate and 500 μL of acetonitrile with 2% formic acid. After centrifugation, the clear supernatant extract (15 μL) was directly injected into the HILIC/MS/MS instrument, and each drug was separated on a Unison UK-Amino column (50 mm × 3 mm i.d., 3 μm particle size) with a linear gradient elution system composed of 10 mM ammonium acetate (pH 6.8) and acetonitrile at a flow rate of 0.7 mL/min. We performed quantification by multiple reaction monitoring (MRM) with negative-ion atmospheric-pressure chemical ionization.
Results:
Distinct peaks were observed for the drugs on each MRM channel within 2 min. The regression equations showed good linearity within the range 0.04-4.0 μg/mL for the tear and plasma samples with detection limits at 0.02-0.04 μg/mL. Recoveries for target analytes (FT and 5-FU) for the tear and plasma samples were in the 94-128% and 94-104% ranges, respectively. The intra- and inter-day coefficients of variation for the two drugs were lower than 10.8%. The accuracies of quantitation were 97-115% for both samples.
Conclusions:
We established a high-throughput, reproducible, and practical procedure for analyzing FT and 5-FU in human tear and plasma samples using HILIC/MS/MS analysis with an aminopropyl-bonded mixed-mode separation column. This method can be applied to the high-throughput routines used in clinical analyses.
More Related Videos
08:56Quantitative Analysis of the Cellular Lipidome of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: March 8, 2020
08:20Author Spotlight: Advancing Tear Fluid Analysis Using a Standardized Protocol for Proteomics Research
Published on: December 1, 2023
Related Concept Videos
Tandem Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: