Continuous thermodilution to assess absolute flow and microvascular resistance: validation in humans using [15O]H2O
Henk Everaars1, Guus A de Waard1, Stefan P Schumacher1
1Department of Cardiology, Amsterdam UMC, Vrije Universiteit, ZH 5F003, De Boelelaan 1117, Amsterdam, the Netherlands.
Aims:
Continuous thermodilution is a novel technique to quantify absolute coronary flow and microvascular resistance (MVR). Notably, intracoronary infusion of saline elicits maximal hyperaemia, obviating the need for adenosine. The primary aim of this study was to validate continuous thermodilution in humans by comparing invasive measurements to [15O]H2O positron emission tomography (PET). As a secondary goal, absolute flow and MVR were compared between invasive measurements obtained with and without adenosine.
Methods And Results:
Twenty-five patients underwent coronary computed tomography angiography (CCTA), [15O]H2O PET, and invasive assessment. Absolute coronary flow and MVR were measured in the left anterior descending and left circumflex artery using a dedicated infusion catheter and a temperature/pressure sensor-tipped guidewire. Invasive measurements were performed with and without adenosine. In order to compare invasive flow measurements with PET perfusion, subtending myocardial mass of the investigated vessels was derived from CCTA using the Voronoi algorithm. Invasive and non-invasive measurements of adenosine-induced hyperaemic flow and MVR showed strong correlation (r = 0.91; P < 0.001 for flow and r = 0.85; P < 0.001 for MVR) and good agreement [intraclass correlation coefficient (ICC) = 0.90; P < 0.001 for flow and ICC = 0.79; P < 0.001 for MVR]. Absolute flow and MVR also correlated well between measurements with and without adenosine (r = 0.97; P < 0.001 for flow and r = 0.98; P < 0.001 for MVR) and showed good agreement (ICC = 0.96; P < 0.001 for flow and ICC = 0.98; P < 0.001 for MVR).
Conclusions:
Continuous thermodilution is an accurate method to measure absolute coronary flow and MVR, which is evidenced by strong agreement with [15O]H2O PET derived flow and resistance. Absolute flow and MVR correlate highly between invasive measurements obtained with and without adenosine, which confirms that intracoronary infusion of room temperature saline elicits steady-state maximal hyperaemia.
More Related Videos
10:20Automation of a Positron-emission Tomography PET Radiotracer Synthesis Protocol for Clinical Production
Published on: October 26, 2018
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Emission Spectra
Reliability and Validity
Mean Absolute Deviation
Let us consider a dataset containing the number of unsold cupcakes in five shops: 10, 15, 8, 7, and 10. Initially, calculate the sample mean. Then calculate the deviation, or the difference, between each data value and the mean. Next, the absolute values of these deviations are added and divided by the sample size to...
Absolute and Local Extreme Values