Mechanical evaluation of the Bravos afterloader system for HDR brachytherapy
Murillo Bellezzo1, José A Baeza2, Robert Voncken2
1Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht, The Netherlands; Centro de Engenharia Nuclear, Instituto de Pesquisas Energéticas e Nucleares IPEN-CNEN/SP, São Paulo, Brazil.
Purpose:
The Bravos afterloader system was released by Varian Medical Systems in October of 2018 for high-dose-rate brachytherapy with 192Ir sources, containing new features such as the CamScale (a new device for daily quality assurance and system recalibration), channel length verification, and different settings for rigid and flexible applicators. This study mechanically evaluated the Bravos system precision and accuracy for clinically relevant scenarios, using dummy sources.
Methods And Materials:
The system was evaluated after three sets of experiments: (1) The CamScale was used to verify inter- and intra-channel dwelling variability and system calibration; (2) A high-speed camera was used to verify the source simulation cable movement inside a transparent quality assurance device, where dwell positions, dwell times, transit times, speed profiles, and accelerations were measured; (3) The source movement inside clinical applicators was captured with an imaging panel while being exposed to an external kV source. Measured and planned dwell positions and times were compared.
Results:
Maximum deviations between planned and measured dwell positions and times for the source cable were 0.4 mm for the CamScale measurements and 0.07 seconds for the high-speed camera measurements. Mean dwell position deviations inside clinical applicators were below 1.2 mm for all applicators except the ring that required an offset correction of 1 mm to achieve a mean deviation of 0.4 mm.
Conclusions:
Features of the Bravos afterloader system provide a robust and precise treatment delivery. All measurements were within manufacturer specifications.
Related Concept Videos
Reaction Mechanisms
For instance, the decomposition of ozone appears to follow a mechanism with two steps:
Mechanical Protein Functions
Mechanism of Conjugation
Mechanical Systems
Compensation Mechanisms
Respiratory Compensation
This mechanism addresses metabolic-induced pH imbalances by adjusting breathing rates. Respiratory compensation begins within minutes of detecting a pH...
An Introduction to Mechanics
According to records, the history of mechanics starts with Aristotle (384–322 BC). He related mechanics to physical theory, aiming for a universal synthesis.
Newton defined mechanics as the branch of physical science that...