Quantitative assessment of the conjunctival microcirculation using a smartphone and slit-lamp biomicroscope
Paul F Brennan1, Andrew J McNeil2, Min Jing3
1Biomedical Sciences Research Institute, Ulster University, Coleraine BT521SA, United Kingdom; Department of Cardiology, Royal Victoria Hospital, Belfast Health and Social Care Trust, Belfast BT126BA, United Kingdom.
Purpose:
The conjunctival microcirculation is a readily-accessible vascular bed for quantitative haemodynamic assessment and has been studied previously using a digital charge-coupled device (CCD). Smartphone video imaging of the conjunctiva, and haemodynamic parameter quantification, represents a novel approach. We report the feasibility of smartphone video acquisition and subsequent haemodynamic measure quantification via semi-automated means.
Methods:
Using an Apple iPhone 6 s and a Topcon SL-D4 slit-lamp biomicroscope, we obtained videos of the conjunctival microcirculation in 4 fields of view per patient, for 17 low cardiovascular risk patients. After image registration and processing, we quantified the diameter, mean axial velocity, mean blood volume flow, and wall shear rate for each vessel studied. Vessels were grouped into quartiles based on their diameter i.e. group 1 (<11 μm), 2 (11-16 μm), 3 (16-22 μm) and 4 (>22 μm).
Results:
From the 17 healthy controls (mean QRISK3 6.6%), we obtained quantifiable haemodynamics from 626 vessel segments. The mean diameter of microvessels, across all sites, was 21.1μm (range 5.8-58 μm). Mean axial velocity was 0.50mm/s (range 0.11-1mm/s) and there was a modestly positive correlation (r 0.322) seen with increasing diameter, best appreciated when comparing group 4 to the remaining groups (p < .0001). Blood volume flow (mean 145.61pl/s, range 7.05-1178.81pl/s) was strongly correlated with increasing diameter (r 0.943, p < .0001) and wall shear rate (mean 157.31 s-1, range 37.37-841.66 s-1) negatively correlated with increasing diameter (r - 0.703, p < .0001).
Conclusions:
We, for the first time, report the successful assessment and quantification of the conjunctival microcirculatory haemodynamics using a smartphone-based system.
Related Concept Videos
Quantitative Analysis
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...
Quantitative Aspects of Drug-Receptor Interaction
Assessment of the Mouth
Mouth Inspection
The inspection begins with visually examining the mouth for symmetry, color, and size.
Nursing Assessment
The nurse collects all aspects of the patient's health in the initial assessment, establishing priorities for ongoing focused assessments...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...
Assessment of the Rectum and Anus
Rectal Inspection
Begin by inspecting the perianal and anal areas for color, texture, rashes,...