Development and validation of a quantitative method for estimation of the urate burden in patients with gouty
Maximilian Kotlyarov1, Kay Geert A Hermann2, Jürgen Mews3
1Department of Radiology (CCM), Charité - Universitätsmedizin Berlin, Campus Mitte, Humboldt-Universität zu Berlin, Freie Universität Berlin, Charitéplatz 1, 10117, Berlin, Germany. maximilian.kotlyarov@charite.de.
Objectives:
To develop a method that allows approximating the mass of monosodium uric acid (MSU) in a gouty tophus using phantom measurements and including tophus density into the calculation of the dual-energy computed tomography (DECT) tophus volumetry.
Methods:
We prepared specimens of different concentrations of MSU placed in an epoxy-based phantom and an excised porcine foreleg. Density and volumetric measurements were performed in sequential single-source DECT scans acquired with increasingly higher tube currents. We developed a method for estimating the tophus mass by multiplying the detected tophus volume with its mean density and adding a specific gravimetric coefficient k. k was derived from the DECT scans by comparing the approximated MSU masses in the epoxy phantom with the known true MSU masses of the specimens.
Results:
Comparison of the approximated MSU masses in the porcine foreleg scans with the true MSU masses of the syringe contents showed similar performance to sole volume measurement while providing additional information on the true uric acid burden: Over 70% of the true urate masses have been detected in MSU concentrations ≥ 45%, while the detection rate was much lower for MSU concentrations ≤ 40%. Retrospective analysis of patients with proven gouty arthritis confirmed the diagnostic potential of the mass approximation technique.
Conclusions:
We successfully established a method to include tophus density measurement for estimation of the uric acid burden in milligrams (instead of ml) in a phantom setting for MSU concentrations above 40%. Future studies should validate its use for follow-up in clinical practice.
Key Points:
• Including tophus density measurement in dual-energy computed tomography scans in a phantom setting can be used for estimating the urate burden in milligrams, which might be useful for imaging follow-up. • The mass [mg] of the uric acid burden in a patient with gouty arthritis can be calculated by multiplying volume [ml] with mean density [HU] using a specific gravimetric coefficient. • Retrospective analysis of two patients with gouty tophi showed the relevance of measuring urate mass in addition to urate volume alone.
More Related Videos
12:09Patient-specific Modeling of the Heart: Estimation of Ventricular Fiber Orientations
Published on: January 8, 2013
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Area Computation by the Alternative Coordinate Method
In Vitro Drug Release Testing: Overview, Development and Validation
Reliability and Validity
Nuclear Binding Energy