Novel 3-dimensionally printed patient-specific guide improves accuracy compared with standard total shoulder
Brandon C Cabarcas1, Gregory L Cvetanovich2, Alejandro A Espinoza-Orias1
1Midwest Orthopaedics at Rush, Rush University Medical Center, Chicago, IL, USA.
Background:
Patient-specific instrumentation (PSI) systems for total shoulder arthroplasty (TSA) can improve glenoid component placement, but may involve considerable expense and production delays. The purpose of this study was to evaluate a novel technique for in-house production of 3-dimensionally printed, patient-specific glenoid guides. We hypothesized that our PSI guide would improve the accuracy of glenoid guide pin placement compared with a standard TSA guide.
Methods:
We randomized 20 cadaveric shoulders to receive pin placement via the PSI guide (n = 10, study group) or standard TSA guide (n = 10, control group). PSI guides were designed to fit each glenoid based on 3-dimensional scapular models constructed from computed tomography scans. A presurgical plan was created for the guide pin trajectory in neutral version and inclination based on individual scapular anatomy. After pin placement, 3-dimensional models from repeated computed tomography scans were superimposed to calculate deviation from the presurgical plan for each specimen.
Results:
Inclination deviation was significantly lower in the PSI group than in the standard guide group (1.5° ± 1.6° vs. 6.4° ± 5.0°, P = .009). The glenoid entry site exhibited significantly less deviation in the PSI group (0.8 ± 0.6 mm vs. 2.1 ± 1.2 mm, P = .008). The average production cost and time for the PSI guides were $29.95 and 4 hours 40 minutes per guide, respectively.
Conclusions:
The PSI guide significantly improved the accuracy of glenoid pin placement compared with the standard TSA guide. Our PSI guides can be produced in-house, inexpensively, and with substantially reduced time compared with commercially available guides.
Related Concept Videos
Improving Translational Accuracy
Improving Translational Accuracy
Muscles of the Shoulder
Anterior Thoracic Muscles
The anterior thoracic muscles include the serratus anterior, subclavius, and...
Uncertainty in Measurement: Accuracy and Precision
Dimensional Analysis
Conversion Factors and Dimensional Analysis
The unit...
Accuracy and Precision