Testing of the Eschweiler 2500-11 blood gas analyser during inflight conditions
D Maillard1, F Ferracci, T Tran Minh
1Laboratoire de Physiologie, Faculté de Médecine Saint-Antoine, Paris, France.
Abstract:
Measurement of blood gas values during air transport of intensive care patients must take into account the extreme inflight variations in pressure. We tested the accuracy and reliability of the Eschweiler 2500-11 apparatus, with its incorporated pressure transducer, on four different gas mixtures equilibrated with fresh blood at two different altitudes simulated in a decompression chamber. Results for the pressure transducer show: 1) concordance of pressure values with those of ground instrumentation; 2) stable response; and 3) absence of hysteresis. Gas measurement electrodes were shown to have linear responses and were accurate for all tested values except very low PCO2 and very high PO2 figures. We conclude that the performance of this device is satisfactory during inflight conditions.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes