Related Experiment Video
Updated: Jan 21, 2026

Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
In vivo reliability of 3D cephalometric landmark determination on magnetic resonance imaging: a feasibility study
Alexander Juerchott1, Christian Freudlsperger2, Sebastian Zingler3
1Department of Neuroradiology, Heidelberg University Hospital, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany.
Objective:
3D cephalometric analysis performed on cone-beam or multi-slice computed tomography (CBCT, MSCT) has superior diagnostic value compared to 2D cephalometry based on radiographs. However, this comes at the expense of increased radiation risks for the predominantly young patients. Magnetic resonance imaging (MRI) has the potential to overcome this diagnostic dilemma but has not been established for 3D cephalometry so far. Since landmark reliability forms the basis for 3D cephalometry, we evaluated the in vivo reliability of established 3D landmarks using MRI.
Materials And Methods:
Sixteen orthodontic patients underwent MRI at 3 Tesla using a 0.5 mm 3D sequence. On each MRI scan, 44 cephalometric landmarks were determined. Image analysis was performed twice by two independent observers. Intra- and inter-rater agreement was assessed by mean measurement errors and intraclass correlation coefficients (ICCs). Measurement errors were calculated as Euclidean distances and distances for x-, y-, and z-coordinates.
Results:
Overall, MRI-based 3D cephalometric landmarks revealed a high reliability comparable to prior in vivo studies using CBCT/MSCT. Intra- and inter-rater ICCs were consistently higher than 0.9. Intra-rater comparisons showed mean measurement differences (ranges) of 0.87 mm (0.41-1.63) for rater I and 0.94 mm (0.49-1.28) for rater II. Average inter-rater difference was 1.10 mm (0.52-2.29). Distinct differences in reliability between the various landmarks were observed, corresponding well with the landmarks' specific shapes.
Conclusions:
The present study demonstrates that MRI enables reliable determination of 3D cephalometric landmarks in vivo.
Clinical Relevance:
This study emphasizes the potential of MRI to perform treatment planning and monitoring in orthodontics as well as oral and maxillofacial surgery without radiation exposure.