Performance of the cone beam computed tomography-based patient positioning system on the Gamma Knife Icon™
Andy Y Xu1, Yi-Fang Wang1, Tony J C Wang1
1Department of Radiation Oncology, Columbia University Irving Medical Center, New York, NY, USA.
Purpose:
Cone beam computed tomography (CBCT) imaging has been implemented on the Leksell Gamma Knife® Icon™ for assessing patient positioning in mask-based Gamma Knife radiosurgery. The purpose of this study was to evaluate the performance of the CBCT-based patient positioning system as a tool for frameless Gamma Knife radiosurgery.
Methods:
Daily quality assurance (QA) CBCT precision test results from a 12-month period were analyzed for the geometric accuracy and the stability of the imager. The performance of the image acquisition module and the image registration algorithm was evaluated using an anthropomorphic head phantom (CIRS Inc., Norfolk, VA) and a XYZR axis manual positioning stage (TOAUTO Inc., Guangdong, China). The head phantom was fixed on a mask adaptor and manually translated in the X, Y, Z directions or rotated around the X, Y, Z axes in the range of ±10 mm or ±10º. A CBCT scan was performed after each manual position setup followed by an image registration to the reference scan. To assess the overall setup uncertainties in fractionated treatment, two cylindrical Presage phantoms (Heuris Inc., Skillman, NJ) of 15 cm diameter and 10 cm height were irradiated with identical prescription dose and shot placement following standard mask-based treatment workflow according to two different fraction schedules: a single fraction treatment of 7.5 Gy and a 5-fraction treatment with 1.5 Gy per fraction.
Results:
The averaged vector deviations of the four marks from their preset values are 0.087, 0.085, 0.095, and 0.079 mm from the 212 daily QA tests. The averaged displacements in the X, Y, Z coordinates and the pitch, yaw, roll angles from the image registration tests are 0.23, 0.27, 0.14, 0.32º, 0.19º, 0.31º from the manual setup. The corresponding maximum differences are 0.41, 0.33, 0.29 mm, 0.45º, 0.31º, and 0.43º, respectively. Compared to the treatment plan using the 2% & 1 mm criteria, the averaged 2D Gamma passing rate is 98.25% for the measured dose distribution from the Presage phantom with 1-fraction irradiation and 95.12% for the 5-fraction irradiation. The averaged Gamma passing rates are 99.53% and 98.16% for the 1-fraction and 5-fraction irradiations using the 2% & 2 mm criteria.
Conclusions:
The CBCT imager and the image registration algorithm can reproduce phantom position with <0.5 mm/0.5º uncertainty. A systematic contribution from the interfraction phantom repositioning procedure was observed in the Gamma analysis over the irradiated volumes of two end-to-end test phantoms.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Position of Equilibrium in Acid-Base Reactions
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...
Position-effect Variegation