Related Experiment Video
Updated: Jan 21, 2026

In Vitro Multiparametric Cellular Analysis by Micro Organic Charge-modulated Field-effect Transistor Arrays
Published on: September 20, 2021
Clinical Evaluation of a Two-dimensional Liquid-Filled Ion chamber Detector Array for Verification of High Modulation
Miljenko Markovic1, Ganesh Narayanasamy2, Sotirios Stathakis1
1Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA.
Introduction:
Clinical evaluation of a two-dimensional (2D) liquid-filled ion chamber detector array used in the verification of highly modulated small beams of stereotactic body radiation therapy (SBRT) has been conducted.
Materials And Methods:
Measurements with the Octavius 1000 SRS (PTW, Freiburg, Germany) detector with 977 liquid-filled ion chambers were compared against EDR2 film and PTW Octavius Seven29. The performance of detector array has been evaluated on ten SBRT patient plans. Dose profiles of individual and composite fields' calculated using Pinnacle3 treatment planning system were compared against measurements with Octavius 1000 SRS detector array, EDR2 film, and Octavius Seven29 detector. Gamma index and profile comparison were used in the evaluation and assessment of the detector's performance.
Results:
The Gamma index measurements show agreement between Pinnacle3 computations and Octavius 1000 SRS array, PTW Octavius Seven29, and EDR2 film for >90% of the points using 2%, 2 mm tolerance criteria. Profiles obtained with the Octavius 1000 SRS were in agreement with the EDR2 film profiles, demonstrating the detector's superior sampling rate.
Conclusions:
The Octavius 1000 SRS is a dosimetrically accurate device to perform quality assurance checks in SBRT treatments. The broad range of measurements performed in this study quantified the dosimetric accuracy of Octavius 1000 SRS detector in the clinical setup of the small fields in radiotherapy.
Related Concept Videos
Self-Evaluation: Self-Enhancement and Self-Verification
High-Performance Liquid Chromatography: Types of Detectors
Strategies of Self-Presentation II: Self-Verification
Common Ion Effect
Precipitation of Ions
The equation that describes the equilibrium between solid calcium carbonate and its solvated ions is:
Ion Channels
Ion channels are specialized integral membrane proteins on the plasma membrane that allow...