Electrical connectors for neural implants: design, state of the art and future challenges of an underestimated
Julia Koch1, Martin Schuettler, Cristian Pasluosta
1Laboratory for Biomedical Microtechnology, Department of Microsystems Engineering, University of Freiburg, Freiburg, Germany.
Abstract:
Technological advances in electrically active implantable devices have increased the complexity of hardware design. In particular, the increasing number of stimulation and recording channels requires innovative approaches for connectors that interface electrodes with the implant circuitry.
Objective:
This work aims to provide a common theoretical ground for implantable connector development with a focus on neural applications.
Approach:
Aspects and experiences from several disciplines are compiled from an engineering perspective to discuss the state of the art of connector solutions. Whenever available, we also present general design guidelines.
Main Results:
Degradation mechanisms, material stability and design rules in terms of biocompatibility and biostability are introduced. Considering contact physics, we address the design and characterization of the contact zone and review contaminants, wear and contact degradation. For high-channel counts and body-like environments, insulation can be even more crucial than the electrical connection itself. Therefore, we also introduce the requirements for electrical insulation to prevent signal loss and distortion and discuss its impact on the practical implementation.
Significance:
A final review is dedicated to the state of the art connector concepts, their mechanical setup, electrical performance and the interface to other implant components. We conclude with an outlook for possible approaches for the future generations of implants.
Related Concept Videos
Factorial Design
Group Design
Neural Regulation
Components of Stress
Interestingly, the hidden cube faces also experience these stresses, equal and...
Components of Language
Determining Electric Field From Electric Potential
In general, regardless of whether the electric field is uniform, it points in the direction of decreasing potential because the force on a positive...