A closer look at the conventional Winston-Lutz test: Analysis in terms of dose
Juan-Francisco Calvo-Ortega1,2, Sandra Moragues-Femenía1,2, Coral Laosa-Bello1,2
1Servicio de Oncología Radioterápica, Hospital Quirónsalud, Barcelona, Spain.
Aim:
To investigate whether the target-isocenter deviations reported by a conventional Winston-Lutz (WL) test actually reflect the shifts of the measured prescription isodose line with respect to the target.
Background:
A conventional WL test uses a metallic ball as a target that aims at several fields. But this test does not report information on the accuracy of the delivery in terms of dose.
Materials And Methods:
A conventional WL test using a metallic pointer as a target (Pointer-WL test) has been recreated in the Eclipse treatment planning system over an acrylic phantom containing a radiochromic film (Dose-WL test). After Dose-WL test delivery, the shift of the 80% prescription isodose line with respect to the target center (d80%-center) was measured using film dosimetry. The Pointer-WL and Dose-WL tests were performed in 10 different sessions. The isocenter deviation reported by the Pointer-WL test was compared to the d80%-center vector, according to the three patient's directions (Left-Right or LR; Anterior-Posterior or AP; and Superior-Inferior or SI).
Results:
The deviations (mean ± SD) found for the Dose-WL tests (LR: 0.5 ± 0.4 mm; AP: 0.5 ± 0.4 mm; SI: 0.6 ± 0.2 mm) were in most cases less than 1 mm, and they were significantly smaller (all p < 0.05) than the maximum deviations reported by the Pointer-WL tests (LR: 1.3 ± 0.3 mm; AP: 1.2 ± 0.4 mm; SI: 1.1 ± 0.3 mm).
Conclusions:
The Dose-WL test described in this study allows estimating the spatial accuracy of the prescription isodose line.
Related Concept Videos
Sign Convention
The normal force acts perpendicular to the beam's cross-section and can...
Dose Size and Dosing Frequency: Determination Methods
Long-term Depression
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Enolate Mechanism Conventions
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...