Detection of bacteria in burn wounds with a novel handheld autofluorescence wound imaging device: a pilot study
Anouk Pijpe1, Yildiz Ozdemir1, Jan C Sinnige1,2
1Burn Center, Red Cross Hospital, PO Box 1074, 1940 EB, Beverwijk, the Netherlands.
Objective:
To compare the detection of bacteria in burn wounds between an bacterial fluorescence imaging device MolecuLight i:X, (Canada), and standard microbiological swabs.
Methods:
Wounds were swabbed three times on one occasion; once with a standard swab, once with a high-fluorescent area swab, indicating a bacterial load >104 colony-forming units (CFU)/gram and a finally with a non-fluorescent (nF) area swab. Proportion agreement of the microbiological results was calculated and the accuracy of the device to detect relevant bacteria was assessed.
Results:
A total of 14 patients with 20 wounds participated in the study. Median post-burn day at sampling time was 21 days. Of the 20 wounds, nine had a positive swab result in either of the three swabs, and 11 showed a highfluorescent area. Overall, positive and negative proportion agreement between standard swab and high-fluorescent swab sample results were 100%. Sensitivity, specificity, positive and negative predictive values of presence of high-fluorescence were 78%, 64%, 64%, and 78%, respectively. For Pseudomonas aeruginosa detection, these results were 100%, 70%, 44% and 100%, respectively.
Conclusion:
The diagnostic accuracy of the bacterial fluorescence imaging device to detect relevant bacteria in burn wounds was moderate and the reliability was equal to standard swabbing. Further research in larger sample sizes and on the relevance of minimal bacterial load and its potential to help with Pseudomonas aeruginosa management is needed.
Related Concept Videos
Phases of Wound Repair
Formation of Blood Clot
In case of deep injuries, trauma to blood vessels results in blood loss. In the meantime, phospholipids released from the ruptured endothelial cellular membrane are converted into arachidonic...
Burn Injuries
The damage results in the death of skin cells, which can lead to a massive loss of fluid. Dehydration, electrolyte imbalance, and renal and circulatory failure follow, which can be fatal. Burn patients are treated with intravenous fluids to offset...
Pilot and Numeric Relaying
Imaging Studies VII: Vascular Imaging
Imaging Studies II: Ultrasonography
Imaging Studies I: Kidney, Ureter, and Bladder Studies