Validation and method comparison of the use of densitometry to quantify monoclonal proteins in canine sera
Adam Dugger Harris1, Emily Rout1, Anne Avery1
1Department of Microbiology, Immunology, and Pathology, College of Veterinary Medicine and Biomedical Science, Colorado State University, Fort Collins, CO, USA.
Background:
Densitometric quantitation using serum protein electrophoresis (SPE) is used to monitor monoclonal proteins (M-proteins) in human patients but has not been validated in the dog. Serum globulin concentrations, species-specific radial immunodiffusion (RID), and ELISAs are currently used in veterinary medicine.
Objective:
We aimed to compare four methods that quantify M-proteins using densitometry and biuret protein (dM-protein) measurements. We also validated the best performing method and compared it with the RID and ELISA methods for measuring canine serum M-protein.
Method:
Serum from six normal dogs and 83 serum samples from 46 dogs with confirmed monoclonal gammopathies were used. A spike and recovery experiment with purified monoclonal IgG and IgM, inter-run and intra-run variability, linearity under dilution, and lower limit of detection were performed. Results of commercial canine RID and ELISA kits for total class-specific immunoglobulin were compared with dM-proteins.
Results:
The corrected perpendicular drop gating method had <20% error for IgG/γ-globulin and IgM/β-globulin M-protein quantifications. Linearity (r > .99), intra-run CV (1.1%-2.3%), and inter-run CVs (2.0%-3.5%) were acceptable. Correlation between the RID and densitometry results ranged from r = .25 to r = .88, depending on the class. The RID result was greater than that of the biuret total protein in 26/63 (41%) IgA cases. A panel of IgG, IgA, and IgM RIDs failed to correctly identify an IgM paraproteinemia in 6/6 (100%) cases. Densitometry was not comparable with any other tested method.
Conclusion:
Densitometric quantitation is a valid technique for measuring M-proteins in the β- and γ-globulin regions. Immunotyping via RID using the tested kit does not appear to detect IgM. Densitometry is recommended for measuring M-proteins in canine patients.
More Related Videos
Related Concept Videos
Reliability and Validity
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Quantifying Work
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...