Effect of implant abutment connection designs, and implant diameters on screw loosening before and after cyclic
Sara Reda Sammour1, Mohamed Maamoun El-Sheikh1, Attiah Aly El-Gendy1
1Prosthodontic department, Faculty of Dentistry, University of Tanta, Egypt.
Aims:
The purpose of this in-vitro study was to evaluate the screw loosening of two different forms of implant abutment connection designs, and two implant diameters by measuring removal torque value (RTV) before and after cyclic loading.
Materials And Methods:
Twenty implant fixtures were divided equally into 2 groups (N=10): group I fixture with conical hybrid connection (CH), and group II fixture with internal hex connection (IH). Each group was divided equally into two subgroups according to implant diameters: subgroup A (3.3mm), and subgroup B (4.2mm). Each fixture was vertically placed in the center of an acrylic resin block. The samples were fixed to the jig, and an implant abutment connected it with a 20 Ncm tightening torque. The samples were subjected to eccentric cyclic loading (at a distance of 5mm) away from center of abutment at 100,000 cycles. A digital torque gauge was used to evaluate screw loosening by measuring RTVs in (Ncm) before and after cyclic loading. The removal torque loss ratio before and after cyclic loading and the removal torque loss ratio between before and after cyclic loading were calculated and analyzed using the SPSS statistical analysis.
Results:
For GI the initial removal torque loss ratio measurement was (14.45±3.18) and decreased significantly after loading, it was (11.47±3.64). For GII the initial removal torque loss ratio measurement was (20.47±4.99) and increased significantly after loading, being (35.35±4.26). There is no significant effect upon screw loosening for two implant diameters.
Conclusion:
Within the limitations of this study, the results suggested that conical hybrid connections showed a better screw stability than an internal hex connection. Therefore, the use of conical implants can be promoted as they have better screw stability compared to other systems.
Related Concept Videos
Design Example: Calculating Safe Diameter for Wind-Exposed Disc
Design of Columns under a Centric Load
Euler's formula is applicable under the assumption that the column is a perfect, straight, homogenous prism, and it is operating...
Design of Columns under an Eccentric Load
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Study Designs in Epidemiology
Observational studies are those where the researcher does not intervene but rather observes natural variations. They include cross-sectional, cohort, and...
Study Design in Statistics
Does aspirin reduce the risk of heart attacks? Is one brand of fertilizer more effective at growing roses than another? Is fatigue as dangerous to a driver as the influence of alcohol? Questions like these are answered using randomized experiments with proper...