Defining a national reference level for intraoperative radiation exposure in urological procedures: FLASH, a
Nick Simson1, Thomas Stonier2, Narin Suleyman3
1Department of Urology, Guy's Hospital, London, UK.
Objectives:
To define reference levels for intraoperative radiation during stent insertion, ureteroscopy (URS), and percutaneous nephrolithotomy (PCNL); to identify variation in radiation exposure between individual hospitals across the UK, between low- and high-volume PCNL centres, and between grade of lead surgeon.
Patients/Subjects And Methods:
In all, 3651 patients were identified retrospectively across 12 UK hospitals over a 1-year period. Radiation exposure was defined in terms of total fluoroscopy time (FT) and dose area product (DAP). The 75th percentiles of median values for each hospital were used to define reference levels for each procedure.
Results:
Reference levels: ureteric stent insertion/replacement (DAP, 2.3 Gy/cm2 ; FT, 49 s); URS (DAP, 2.8 Gy/cm2 ; FT, 57 s); PCNL (DAP, 24.1 Gy/cm2 ; FT, 431 s). Significant variations in the median DAP and FT were identified between individual centres for all procedures (P < 0.001). For PCNL, there was a statistically significant difference between DAP for low- (<50 cases/annum) and high-volume centres (>50 cases/annum), at a median DAP of 15.0 Gy/cm2 vs 4.2 Gy/cm2 (P < 0.001). For stent procedures, the median DAP and FT differed significantly between grade of lead surgeon: Consultant (DAP, 2.17 Gy/cm2 ; FT, 41 s) vs Registrar (DAP, 1.38 Gy/cm2 ; FT, 26 s; P < 0.001).
Conclusion:
This multicentre study is the largest of its kind. It provides the first national reference level to guide fluoroscopy use in urological procedures, thereby adding a quantitative and objective value to complement the principles of keeping radiation exposure 'as low as reasonably achievable'. This snapshot of real-time data shows significant variation around the country, as well as significant differences between low- and high-volume centres for PCNL, and grade of lead surgeon for stent procedures.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
10:03Characterization of Electrode Materials for Lithium Ion and Sodium Ion Batteries Using Synchrotron Radiation Techniques
Published on: November 11, 2013
Related Concept Videos
Urologic Endoscopic Procedure: Cystoscopic Examination
Biological Effects of Radiation
National Nursing Organizations I
National Nursing Organizations II
The AACN emphasizes a healthy work environment through six standards to achieve an optimal patient outcome. The standards are appropriate staffing, meaningful recognition, collaboration, authentic leadership, effective communication, and decision-making. In addition, AACN provides certification programs, webinars, journals, and...
Inertial Frames of Reference
Non-inertial Frames of Reference