Related Experiment Video
Updated: Jan 20, 2026

Substernal Thyroid Biopsy Using Endobronchial Ultrasound-guided Transbronchial Needle Aspiration
Published on: November 10, 2014
Robotic needle insertion during computed tomography fluoroscopy-guided biopsy: prospective first-in-human feasibility
Takao Hiraki1, Tetsushi Kamegawa2, Takayuki Matsuno3
1Department of Radiology, Okayama University Medical School, 2-5-1 Shikatacho, Kitaku, Okayama, 700-8558, Japan. hiraki-t@cc.okayama-u.ac.jp.
Introduction:
This was a prospective, first-in-human trial to evaluate the feasibility and safety of insertion of biopsy introducer needles with our robot during CT fluoroscopy-guided biopsy in humans.
Materials And Methods:
Eligible patients were adults with a lesion ≥ 10 mm in an extremity or the trunk requiring pathological diagnosis with CT fluoroscopy-guided biopsy. Patients in whom at-risk structures were located within 10 mm of the scheduled needle tract were excluded. Ten patients (4 females and 6 males; mean [range] age, 72 [52-87] years) with lesions (mean [range] maximum diameter, 28 [14-52] mm) in the kidney (n = 4), lung (n = 3), mediastinum (n = 1), adrenal gland (n = 1), and muscle (n = 1) were enrolled. The biopsy procedure involved robotic insertion of a biopsy introducer needle followed by manual acquisition of specimens using a biopsy needle. The patients were followed up for 14 days. Feasibility was defined as the distance of ≤ 10 mm between needle tip after insertion and the nearest lesion edge on the CT fluoroscopic images. The safety of robotic insertion was evaluated on the basis of machine-related troubles and adverse events according to the Clavien-Dindo classification.
Results:
Robotic insertion of the introducer needle was feasible in all patients, enabling pathological diagnosis. There was no machine-related trouble. A total of 11 adverse events occurred in 8 patients, including 10 grade I events and 1 grade IIIa event.
Conclusion:
Insertion of biopsy introducer needles with our robot was feasible at several locations in the human body.
Key Points:
• Insertion of biopsy introducer needles with our robot during CT fluoroscopy-guided biopsy was feasible at several locations in the human body.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Clinical Trials
There are four phases in a clinical trial. A phase one...
Clinical Trials: Overview
Trial and Error and Algorithm
Design Example: Traverse Angle Computations