Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
The challenging right inferior pulmonary vein: A systematic approach for successful cryoballoon ablation
Raphaël P Martins1, Amélie Nicolas1, Vincent Galand1
1Inserm, LTSI-UMR 1099, service de cardiologie et maladies vasculaires, Univ Rennes, CHU de Rennes, 2, rue Henri-Le-Guilloux, 35000 Rennes, France.
Background:
Pulmonary vein isolation (PVI) using cryoballoon ablation is widely used for rhythm control in patients with paroxysmal atrial fibrillation. This technique has a steep learning curve, and PVI can be achieved quickly in most patients. However, the right inferior pulmonary vein (RIPV) is often challenging to occlude and isolate.
Aim:
We aimed to analyse the efficacy of RIPV ablation using a systematic approach.
Methods:
Consecutive patients referred for cryoballoon ablation of paroxysmal atrial fibrillation were enrolled prospectively. A systematic approach was used for RIPV cryoablation. The primary endpoint was acute RIPV isolation during initial freeze.
Results:
A total of 214 patients were included. RIPV isolation during initial freeze occurred in 179 patients (82.2%). Real-time PVI could be observed in 72 patients (33.6%), whereas cryoballoon stability required pushing the Achieve™ catheter inside the RIPVs in the remaining patients. The rate of unsuccessful or aborted first freeze as a result of insufficient minimal temperature was significantly higher in patients with real-time pulmonary vein potential recording (16.7% vs. 6.3%; P=0.031). To overcome this issue and obtain both stability and real-time PVI, a dedicated "whip technique" was developed. Twelve patients (5.6%) required a redo ablation; only two of these had a reconnected RIPV.
Conclusions:
A systematic approach to RIPV cryoablation can lead to a high rate of first freeze application. Operators should not struggle to visualize pulmonary vein potentials before ablation, as this may decrease cryoapplication efficacy. Thus, stability should be preferred over real-time PVI for RIPV ablation. Both stability and real-time PVI can be obtained using a "whip technique".
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
05:22Author Spotlight: Demonstrating Systematic Endobronchial Ultrasound to New Endoscopists
Published on: August 11, 2023
Related Concept Videos
Chemical Equilibria: Systematic Approach to Equilibrium Calculations
The first step is to identify all the chemical reactions involved, The...
Ecological Succession
Random and Systematic Errors
Systematic Sampling Method
Systematic sampling is one of the simplest methods...
Veins
Structure of Veins:
The structure of veins is specifically designed to assist in the low-pressure transportation of...
Propagation of Uncertainty from Systematic Error