High resolution radiochromic film dosimetry: Comparison of a microdensitometer and an optical microscope
P Pellicioli1, S Bartzsch2, M Donzelli3
1The European Synchrotron Radiation Facility, ID17 Biomedical Beamline, Grenoble, France; Inserm UA7 STROBE, Grenoble Alpes University, Grenoble, France; Swansea University Medical School, Singleton Park, Swansea SA2 8PP, United Kingdom.
Purpose:
Microbeam radiation therapy is a developing technique that promises superior tumour control and better normal tissue tolerance using spatially fractionated X-ray beams only tens of micrometres wide. Radiochromic film dosimetry at micrometric scale was performed using a microdensitometer, but this instrument presents limitations in accuracy and precision, therefore the use of a microscope is suggested as alternative. The detailed procedures developed to use the two devices are reported allowing a comparison.
Methods:
Films were irradiated with single microbeams and with arrays of 50 µm wide microbeams spaced by a 400 µm pitch, using a polychromatic beam with mean energy of 100 keV. The film dose measurements were performed using two independent instruments: a microdensitometer (MDM) and an optical microscope (OM).
Results:
The mean values of the absolute dose measured with the two instruments differ by less than 5% but the OM provides reproducibility with a standard deviation of 1.2% compared to up to 7% for the MDM. The resolution of the OM was determined to be ~ 1 to 2 µm in both planar directions able to resolve pencil beams irradiation, while the MDM reaches at the best 20 µm resolution along scanning direction. The uncertainties related to the data acquisition are 2.5-3% for the OM and 9-15% for the MDM.
Conclusion:
The comparison between the two devices validates that the OM provides equivalent results to the MDM with better precision, reproducibility and resolution. In addition, the possibility to study dose distributions in two-dimensions over wider areas definitely sanctions the OM as substitute of the MDM.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits
Molecular Comparison of Gases, Liquids, and Solids