Iterative reconstruction for image enhancement and dose reduction in diagnostic cone beam CT imaging
Dmitri Matenine1,2, Matthieu Schmittbuhl1,3, Stéphane Bedwani4,5
1Laboratoire de recherche en imagerie et orthopédie, Centre de recherche du Centre hospitalier de l'Université de Montréal, Montréal, QC, Canada.
Background:
Iterative reconstruction is well-established in diagnostic multidetector computed tomography (MDCT) for dose reduction and image quality enhancement. Its application to diagnostic cone beam computed tomography (CBCT) is only emerging and warrants a quantitative evaluation.
Methods:
Several phantoms and a canine head specimen were imaged using a commercially available small-field CBCT scanner. Raw projection data were reconstructed using the Feldkamp-Davis-Kress (FDK) method with different filters, including denoising via total variation (TV) minimization (FDK-TV). Iterative reconstruction was carried out using the TV-regularized ordered subsets convex technique (OSC-TV). Signal-to-noise ratio (SNR), noise power spectrum (NPS) and spatial resolution of images were estimated. Dose levels were measured via the weighted computed tomography dose index, while low-dose image quality degradation was estimated via structural similarity (SSIM).
Results:
OSC-TV and FDK-TV were shown to significantly improve image signal-to-noise ratio (SNR) compared to FDK with a standard filter, 5.8 and 4.0 times, respectively. Spatial resolution attained with different algorithms varied moderately across different experiments. For low-dose acquisitions, image quality decreased dramatically for FDK but not for FDK-TV nor OSC-TV. For low-dose canine head images acquired using about 1/5 of the dose compared to a reference image, SSIM dropped to about 0.3 for FDK, while remaining at 0.92 for FDK-TV and 0.96 for OSC-TV.
Conclusion:
OSC-TV was shown to improve image quality compared to FDK and FDK-TV. Moreover, this iterative approach allowed for significant dose reduction while maintaining image quality.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Oxidation-Reduction Reactions
Dose Size and Dosing Frequency: Determination Methods