Related Experiment Video
Updated: Jan 20, 2026

Assessment of Physical Activity Intensity with Accelerometers and Oxygen Consumption
Published on: June 20, 2025
Field-based Measurement of Sleep: Agreement between Six Commercial Activity Monitors and a Validated Accelerometer
Andrew G Kubala1, Bethany Barone Gibbs1, Daniel J Buysse2
1Department of Health and Physical Activity, University of Pittsburgh , Pittsburgh, Pennsylvania.
Objective:
To examine agreement between multiple commercial activity monitors (CAMs) and a validated actigraph to measure sleep.
Methods:
Thirty adults without sleep disorders wore an Actiwatch Spectrum (AW) and alternated wearing 6 CAMs for one 24-h period each (Fitbit Alta, Jawbone Up3, Misfit Shine 2, Polar A360, Samsung Gear Fit2, Xiaomi Mi Band 2). Total sleep time (TST) and wake after sleep onset (WASO) were compared between edited AW and unedited CAM outputs. Comparisons between AW and CAM data were made via paired t-tests, mean absolute percent error (MAPE) calculations, and intra-class correlations (ICC). Intra-model reliability was performed in 10 participants who wore a pair of each AW and CAM model.
Results:
Fitbit, Jawbone, Misfit, and Xiaomi overestimated TST relative to AW (53.7-80.4 min, P ≤ .001). WASO was underestimated by Fitbit, Misfit, Samsung and Xiaomi devices (15.0-27.9 min; P ≤ .004) and overestimated by Polar (27.7 min, P ≤ .001). MAPEs ranged from 5.1% (Samsung) to 25.4% (Misfit) for TST and from 36.6% (Fitbit) to 165.1% (Polar) for WASO. TST ICCs ranged from .00 (Polar) to .92 (Samsung), while WASO ICCs ranged from .38 (Misfit) to .69 (Samsung). Differences were similar between poor sleepers (Pittsburgh Sleep Quality Index global score >5; n = 10) and good sleepers. Intra-model reliability analyses revealed minimal between-pair differences and high ICCs.
Conclusions:
Agreement between CAMs and AW varied by device, with greater agreement observed for TST than WASO. While reliable, variability in agreement across CAMs with traditional actigraphy may complicate the interpretation of CAM data obtained for clinical or research purposes.
More Related Videos
Related Concept Videos
Reliability and Validity
Insufficient Sleep and Sleep Deprivation
Sleep deprivation is a more severe form of sleep loss...
Stages of Sleep
Before sleep begins, in wakefulness, the brain exhibits primarily beta waves, which are high in frequency and low in amplitude, indicating alertness...
Understanding Sleep
The circadian rhythm, a nearly 24-hour cycle, is deeply influenced by environmental light cues. Light exposure directly affects the hypothalamus, which in turn regulates...
Sleep Apnea
The condition is more prevalent among...
Techniques of therapeutic communication I: Active Listening, Sharing Observations, Validation, and Using Touch
Therapeutic communication is not the same as social interaction. Social interaction has no goal or purpose and consists of casual information sharing, whereas therapeutic communication has a plan or purpose for the conversation. Therapeutic...