Quantitative assessment of variation in CBCT image technical parameters related to CBCT detector lateral-offset position
Related Experiment Video
Updated: Jan 20, 2026

Wideband Optical Detector of Ultrasound for Medical Imaging Applications
Published on: May 11, 2014
Gustavo Machado Santaella1, Pedro Luiz Rosalen2, Polyane Mazucatto Queiroz1
1Department of Oral Diagnosis, Area of Oral Radiology, Piracicaba Dental School, University of Campinas, Piracicaba, Sao Paulo, Brazil.
Objectives:
To assess the effect of CBCT detector position (aligned/lateral-offset) on image technical parameters (mean voxel value - MVV, standard deviation of voxel value (SDVV) distribution), comparing peripheral regions of interest (ROIs) to the central ROI in CBCT volumes.
Methods:
40 CBCT volumes of a wax phantom were acquired in six units with aligned and/or lateral-offset detectors: Cranex 3Dx (CRA), Ortophos SL (ORT), Picasso Trio (PIC), Promax 3D Mid (PRO), Scanora 3D (SCA), and X1. Four image-acquisition protocols used an aligned detector, and four a lateral-offset detector. In each volume, 13 ROIs (12 peripheral and 1 central) were evaluated. MVV and SDVV of the peripheral ROI were compared to those of the central ROI in the volume. MVVD (the difference in percentage, between the MVV of a peripheral and the central ROI) was calculated.
Results:
For aligned-detectors, MVV increased (ORT and PRO) or decreased (CRA and X1) in the ROIs farther from the centre. For lateral-offset detectors, ROIs farther from the centre showed increased MVV. SDVV for most aligned detectors was lower, the nearer the ROI was to the centre. For lateral-offset detectors, it was lower for the peripheral ROIs, except with PIC. Range for MVVD was -32.8% to 22.8% for units with aligned detectors, and -20.7% to 69.5% for lateral-offset detectors.
Conclusion:
Lateral-offset detectors to acquire CBCT images significantly change SDVV distribution within the field-of-view, and lead to MVVD with increased range, compared to aligned detectors. This must be taken in consideration in the clinic, if voxel-value dependent measurements are to be performed.
Related Concept Videos
08:21Wideband Optical Detector of Ultrasound for Medical Imaging Applications
06:18Pedicle Screw Placement Using an Augmented Reality Head-Mounted Display in a Porcine Model
09:39Quantitative Immunofluorescence Assay to Measure the Variation in Protein Levels at Centrosomes
06:35Ultrasound Assessment of Endothelial Function: A Technical Guideline of the Flow-mediated Dilation Test
05:49Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
06:48Three-Dimensional Echocardiographic Method for the Visualization and Assessment of Specific Parameters of the Pulmonary Veins