Related Experiment Video
Updated: Jan 20, 2026

Author Spotlight: Integrating Computational and Experimental Approaches in Precision Oncology
Published on: December 1, 2023
Initial experience with vascular plug devices for mechanical thrombectomy in symptomatic neonates and infants
Eimear McGovern1, Athar M Qureshi2, Bryan H Goldstein1
1Department of Pediatrics, University of Cincinnati College of Medicine, The Heart Institute, Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio.
Objectives:
To describe a novel technique for mechanical thrombectomy (MT) in young children.
Background:
Acute thrombosis in children may be a cause of significant morbidity or mortality. MT options in children are limited due to patient and vessel size, prompting consideration of innovative approaches.
Methods:
Two-center review of children with symptomatic non-cerebral thrombotic vessel occlusion. A novel approach to MT was performed using a deployed but unreleased amplatzer vascular plug (AVP) to "scrape" the target vessel in retrograde fashion, thereby harvesting the thrombus via the access sheath.
Results:
AVP-MT was performed in four patients at a median age of 2 months (range: 1 day-11 months) and median weight of 5.6 kg (1.8-10.9). In two cases, AVP-MT was performed in isolation in neonates on cervical extracorporeal membrane oxygenation (ECMO) support with cannula-associated obstructive aortic and brachiocephalic thrombus. AVP-MT was also performed as adjunctive therapy in a patient with single ventricle physiology and pulmonary artery thrombosis and in an infant with mitral valve endocarditis and femoral arterial thromboembolism. In all cases, AVP-MT was used to successfully remove thrombus from the target vessel and restore arterial flow, without procedural complications. Recurrent thrombosis occurred in both neonates, due to persistence of the ECMO cannula, with subsequent mortality. The other children were free of recurrent thrombus at 2 and 34 months.
Conclusions:
Acute relief of symptomatic thrombosis may be achieved using AVP-MT in small children, although thrombosis frequently recurs, and clinical prognosis may remain poor. This strategy should be considered when alternative MT treatment options are not applicable.
Related Concept Videos
07:03Bilayer Microfluidic Device for Combinatorial Plug Production
Stromal Vascular Fraction-Enriched Fat Grafting for the Management of Symptomatic Neuromas
07:27How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
11:50Clinical Practice Protocol of Creative Music Therapy for Preterm Infants and Their Parents in the Neonatal Intensive Care Unit
11:39Non-invasive Optical Measurement of Cerebral Metabolism and Hemodynamics in Infants
08:45New Thrombectomy Technique for Total Portal Vein Thrombosis in Liver Transplantation