Accuracy of Glucose Meter among Adults in a Semi-urban Area in Kathmandu, Nepal
Asmita Pokhrel1, Vinutha Silvanus2, Buddhi Raj Pokhrel3
1Department of Biochemistry, Nepal Medical College and Teaching Hospital, Kathmandu, Nepal.
Introduction:
Glucose meters are gaining popularity in monitoring of blood glucose at household levels and in health care set-ups due to their portability, affordability and convenience of use over the laboratory based reference methods. Still they are not free of limitations. Operator's technique, extreme temperatures, humidity, patients' medication, hematocrit values can affect the reliability of glucose meter results. Hence, the accuracy of glucose meter has been the topic of concern since years. Therefore, present study aims to evaluate the analytical and clinical accuracy of glucose meter using International Organization for Standardization 15197 guideline.
Methods:
A community based descriptive cross-sectional study was conducted in Kapan, Kathmandu, Nepal in April 2018. Glucose levels were measured using glucose meter and reference laboratory method simultaneously among 203 adults ≥20 years, after an overnight fasting and two hours of ingestion of 75 grams glucose. Modified Bland-Altman plots were created by incorporating ISO 15197 guidelines to check the analytical accuracy and Park error grid was used to evaluate the clinical accuracy of the device.
Results:
Modified Bland-Altman plots showed>95% of the test results were beyond the acceptable analytical criteria of ISO 15197:2003 and 2013. Park Error Grid-Analysis showed 99% of the data within zones A and B of the consensus error grid.
Conclusions:
Glucose meter readings were within clinically acceptable parameters despite discrepancies on analytical merit. Possible sources of interferences must be avoided during the measurement to minimize the disparities and the values should be interpreted with caution.
Related Concept Videos
Using a pH Meter
Acids and bases are substances capable of donating protons (H+) and hydroxide ions (OH-), respectively. They are two extremes that describe chemicals. Mixing acids and bases can cancel out or neutralize their extreme effects. A substance that is neither acidic nor basic is neutral. The values of proton concentration ([H+]) for most solutions are inconveniently small and difficult to compare so that a more practical...
13:48Design and Construction of an Urban Runoff Research Facility
Using GIS to Investigate Urban Forestry
Urban forests broadly include urban parks, street trees, landscaped boulevards, public gardens, river and coastal promenades, greenways, river corridors, wetlands, nature preserves, natural areas, shelterbelts of trees, and working trees at industrial brownfield sites. The history of urban trees begins with trees as landscape embellishment. Today, urban trees are seen as essential components of city infrastructure...
14:55Evaluating the Effect of Roadside Parking on a Dual-Direction Urban Street
08:49Vegetated Treatment Systems for Removing Contaminants Associated with Surface Water Toxicity in Agriculture and Urban Runoff
Improving Translational Accuracy