System accuracy assessments with a blood glucose meter with combined glucose and ß-hydroxybutyrate measurement
Sylvia Weissenbacher1, Chen-Yu Yang2, Tang-Chin Kuan3
1MED TRUST , Marz , Austria.
Abstract:
Background: We evaluated the Wellion Galileo GLU/KET blood and ketone (ß-Hydroxybutyrate, ß-OHB) meter to demonstrate that it meets ISO15107:2015 regulatory approval criteria. Research Design and Methods: A total of 100 subjects (52 female, age: 30 to 84 years, diabetes: 10 type 1/90 type 2) with blood glucose levels distributed over the entire measurement range as required by the ISO15197 protocol were tested (double determinations with 3 strip lots and two devices). A similar test protocol was followed to test ß-OHB strip performance (reference devices: YSI 2300plus for glucose and STANBIO ß-HOB LiquiColor TestKit for ß-OHB). Precision was tested for glucose with 3 blood glucose concentrations (ß-OHB: 2 control solutions). Results: All glucose test-strip lots met the strict ISO acceptance criteria. Mean absolute relative difference (MARD) was 4.9% and all data pairs were in zone A of the consensus error grid. The ß-OHB test-strips also met the pre-defined acceptance criteria. Within-run and between-run precision was calculated to be 2.3% and 0.7% for the glucose strips (3.7%/0.8% for the ketone strips). Conclusions: When tested according to the ISO15197:2015 guideline, the device showed very accurate measurement performance for glucose and ß-OHB testing and fully met regulatory accuracy approval criteria.
More Related Videos
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
10:35Extracellular Glucose Depletion as an Indirect Measure of Glucose Uptake in Cells and Tissues Ex Vivo
Published on: April 6, 2022
Related Concept Videos
06:28A Simple Flow Cytometric Method to Measure Glucose Uptake and Glucose Transporter Expression for Monocyte Subpopulations in Whole Blood
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
02:23Oral Glucose Tolerance Test (OGTT): Measuring Glucose Metabolism in Mice
10:35Extracellular Glucose Depletion as an Indirect Measure of Glucose Uptake in Cells and Tissues Ex Vivo
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
08:01A Method for Manipulating Blood Glucose and Measuring Resulting Changes in Cognitive Accessibility of Target Stimuli