Reducing temperature errors in transcranial MR-guided focused ultrasound using a reduced-field-of-view sequence
William A Grissom1,2, Steven Allen3
1Vanderbilt University Institute of Imaging Science, Nashville, Tennessee.
Purpose:
To reduce temperature errors due to water motion in transcranial MR-guided focused ultrasound (tcMRgFUS) ablation.
Theory And Methods:
In tcMRgFUS, water is circulated in the transducer bowl around the patient's head for acoustic coupling and heat removal. The water moves during sonications that are monitored by MR thermometry, which causes it to alias into the brain and create temperature errors. To reduce these errors, a two-dimensional excitation pulse was implemented in a gradient-recalled echo thermometry sequence. The pulse suppresses water signal by selectively exciting the brain only, which reduces the imaging FOV. Improvements in temperature precision compared to the conventional full-FOV scan were evaluated in healthy subject scans outside the tcMRgFUS system, gel phantom scans in the system with heating, and in 2×-accelerated head phantom scans in the system without heating.
Results:
In vivo temperature precision (standard deviation of temperature errors) outside the tcMRgFUS system was improved 43% on average, due to the longer TR and TE of the reduced-FOV sequence. In the phantom heating experiments, the hot spot was less distorted in the reduced-FOV scans, and background temperature precision was improved 59% on average. In the accelerated head phantom temperature reconstructions, temperature precision was improved 89% using the reduced-FOV sequence.
Conclusions:
Reduced-FOV temperature imaging alleviates temperature errors due to water bath motion in tcMRgFUS, and enables accelerated temperature mapping with greater precision.
More Related Videos
05:54MRI-guided Focused Ultrasound Thalamotomy for Patients with Medically-refractory Essential Tremor
Published on: December 13, 2017
08:08Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
Related Concept Videos
07:07Errors as a Means of Reducing Impulsive Food Choice
05:54MRI-guided Focused Ultrasound Thalamotomy for Patients with Medically-refractory Essential Tremor
08:08Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Reducing Agents
Controlling the reactivity and selectivity during the synthesis of a molecule is very important criteria for chemists. This has led to the development of many reagents that allow chemists to pick and choose reagents suitable for a given task. Quite often, a balance between reactivity and selectivity needs to be achieved. This experiment will use IR spectroscopy to monitor the reaction and to...
Reducing Line Loss
With a step-up transformer at the source, the voltage is increased, thereby reducing the current in the transmission lines since power loss in...
09:57Obtaining Quality Extended Field-of-View Ultrasound Images of Skeletal Muscle to Measure Muscle Fascicle Length