Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Performance of Dynamic Automated CT Annular Measurements Compared to Standard Manual Measurements for Transcatheter
Quynh A Truong1,2, Alan C Legasto1, Roderick C Deaño3
1Department of Radiology, Weill Cornell Medicine, New York, NY.
Purpose:
We sought to determine the performance of an automated computed tomography (CT) software that provides dynamic annular measurements of all available cardiac phases for transcatheter aortic valve replacement (TAVR) sizing as compared to the standard single manual measurement.
Materials And Methods:
In 60 TAVR patients (84±7 years, 60% male) who underwent pre-procedural CT scans, we measured the aortic annular diameters, perimeter, and area using (1) the dynamic automated CT measurements and (2) standard single manual measurement from the cardiac phase of maximum systolic opening by visual estimate.
Results:
The automated software was successful in providing annular measurements in 43/60 (72%) of cases, with the remainder requiring semi-automated contours. The maximum dynamic automated values were predominantly in systole (46/60[77%] for diameter, 44/60[73%] for perimeter, 48/60[80%] for area), and was a different phase from the standard manual phase in 46/60 (77%) cases. The maximum dynamic automated annular values were larger than the standard manual values measured (Δdiameter 0.35 mm, p=0.04; Δperimeter 1.71 mm, p<0.001; Δarea 15.6 mm2, p<0.001). When comparing standard manual to the same phase by automated measurements, while there was no difference in annular mean diameter (p=0.80), perimeter and area were larger with the automated measurements (Δperimeter 0.95 mm, p=0.002; Δarea 10.8 mm2, p=0.03). However, the maximum automated measurements were consistently larger than the same phase automated measurements (Δdiameter 0.13 mm, p<0.001; Δperimeter 0.42 mm, p<0.001; Δarea 4.4 mm2, p<0.001).
Conclusions:
Automated maximum dynamic CT annular measurements provide larger values than standard manual and same phase automated measurements.
More Related Videos
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
07:46Automatic Surgery in Transcatheter Aortic Valve Replacement Using Augmented Reality
12:17Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots