Does an iPad fixation disparity test give equivalent results to the Mallett near fixation disparity test?
Ketan R Parmar1, Christine Dickinson1, Bruce J W Evans2
1Division of Pharmacy and Optometry, School of Health Sciences, Faculty of Biology, Medicine and Health, University of Manchester, Manchester Academic Health Science Centre, Manchester, UK.
Background:
Various instruments have been developed to measure aligning prism, the prism that eliminates a fixation disparity (associated heterophoria). This includes the established Mallett near vision unit and recently developed Thomson Vision Toolbox on the iPad. With no previous research investigating the agreement between these instruments, practitioners may question if they can be used interchangeably.
Methods:
80 participants underwent near vision testing with the Mallett unit and iPad fixation disparity test. Data were analysed in four ways to investigate the agreement of the two instruments.
Results:
Many participants reported no fixation disparity (horizontally 46.25%, vertically 82.5%), or non-significant aligning prism (horizontally 70%, vertically 97.5%), on both instruments. The iPad revealed a larger range of aligning prism results horizontally, 6Δ base out to 15Δ base in; the Mallett unit produced a larger range of results vertically, 1Δ base up to 3.5Δ base down. More participants required a significant aligning prism on the Mallett unit horizontally and vertically. Wilcoxon signed rank analysis found that the difference in aligning prism was not statistically significant (p=0.357 horizontally, p=0.236 vertically), but 95% limits of agreement revealed clinically significant differences between the instruments.
Conclusion:
Although the measured differences between the instruments are not significant in a Wilcoxon analysis, a Bland & Altman approach shows them to be in some cases clinically unacceptable, therefore the instruments should not be used interchangeably. Previous research indicates that the Mallett unit performs reasonably well at detecting symptomatic individuals and determining a prismatic correction that is likely to be helpful. Further research is required to determine the performance of the iPad test in these functions and to assess the reproducibility of both instruments.
Related Concept Videos
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Carbon-dioxide Fixation
10:21Protocol for HER2 FISH Using a Non-cross-linking, Formalin-free Tissue Fixative to Combine Advantages of Cryo-preservation and Formalin Fixation
05:22Iris Fixation via External Pentagram Suturing
06:57Instillation and Fixation Methods Useful in Mouse Lung Cancer Research
05:48Lung Fixation under Constant Pressure for Evaluation of Emphysema in Mice