Minimal defibrillation thresholds and the correlation with implant position in subcutaneous implantable-defibrillator
Anne-Floor B E Quast1, Sarah W E Baalman1, Willeke Van der Stuijt1
1Department of Clinical and Experimental Cardiology, Heart Center, Amsterdam Cardiovascular Sciences, Amsterdam UMC, University of Amsterdam, Amsterdam, The Netherlands.
Background:
Since the introduction of the subcutaneous implantable-defibrillator (S-ICD) knowledge of factors elevating the defibrillation threshold (DFT), have increased. Optimal device positioning most likely results in a decrease in DFT. The PRAETORIAN score is a tool to systematically evaluate S-ICD implant position and predict conversion success by estimating the DFT on a chest X-ray. The objective of this study is to determine DFT in de novo S-ICD patients.
Methods:
De novo S-ICD patients were enrolled with DFT testing using a single 30 J shock or a prespecified step-down protocol. Chest X-rays were obtained and implant position was evaluated using the PRAETORIAN score.
Results:
Fifteen patients, age 47 (±18) years and body mass index (BMI) 25(±3) kg/m2 , underwent conversion testing with a single 30 J shock. Fourteen patients (93%) converted successfully. Implant position was optimal in all patients, PRAETORIAN score of 30 (100%). Twelve patients were available for analysis in the step-down DFT cohort, age 46 years (±15), left ventricular ejection fraction 50% (±15%) and a median BMI of 25.7 kg/m2 (range 23.8-29.1 kg/m2 ). No complications occurred during DFT testing. Mean DFT was 29 J (±12 J). Time to shock was 11 seconds (±2 seconds) for a 20 J shock compared with 17 seconds in case of a 70 J shock output. Nearly all patients were implanted with optimal device positioning with the lowest possible PRAETORIAN score of 30 (92%).
Conclusion:
DFT in S-ICD patients with optimal device positioning is lower than previously reported. Conversion testing using a low shock output reduced time to therapy by 6 seconds on average.
Related Concept Videos
05:51Computed Tomography (CT) Guided Implantation of a Totally Implantable Venous Access Port (TIVAP) through Subclavian Vein
16:11Implantation of the Syncardia Total Artificial Heart
08:153D Planning and Printing of Patient Specific Implants for Reconstruction of Bony Defects
03:44A Unique Mouse Model for Quantitative Assessment of Biofilm Formation on Surgical Implants in Subcutaneous Abscess
09:56Cochlear Implantation in the Guinea Pig
08:38Renal Capsule Xenografting and Subcutaneous Pellet Implantation for the Evaluation of Prostate Carcinogenesis and Benign Prostatic Hyperplasia