Global multielectrode contact-mapping plus ablation with a single catheter in patients with atrial fibrillation:
Hans Kottkamp1, Gerhard Hindricks2, Christian Pönisch1
1Department of Electrophysiology, Hirslanden Hospital, Zurich, Switzerland.
Introduction:
The critical question for technological advancement of catheter ablation of atrial fibrillation (AF) is whether a creative new concept can combine and even improve the options of single-tip catheters with the simplicity of the use of balloon catheters. Herein are described the results from the first clinical study of a new multielectrode contact-mapping plus ablation array (Globe) offering such a complete solution.
Methods And Results:
The multielectrode Globe array consists of 16 flat ribs with 122 gold-plated electrodes. Each electrode can record electrograms, ablate, pace, and can measure tissue contact and temperature. Single-shot pulmonary vein isolation (PVI) is possible with temperature-guided ablation of up to 24 electrodes simultaneously with automatic, individual power control of every electrode. Sixty patients with symptomatic AF underwent PVI using the Globe. In all sixty patients, acute PVI was achieved in 232 of 234 attempted PVs (99.1%). In 34 patients treated with "single-hot-shot" ablation, PVI was achieved in 136 of 136 PVs (100%). Single-procedure 12-month freedom from AF off antiarrhythmic drugs in the "single-hot-shot" group was 75.5% and freedom from AF/atrial tachycardia 72.3%. In two patients, pericardial tamponade was observed, one after a transseptal puncture, and one during array insertion with an over-advanced sheath. There were no other device-related serious adverse events, including stroke, PV stenosis, esophageal perforation, or phrenic nerve palsy.
Conclusions:
In this first clinical series, the Globe catheter was found to be an easy-to-use system for single-shot PVI. The continuously updated multielectrode voltage and activation mapping data indicate future options for mapping and ablation beyond PVI.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
Related Concept Videos
11:21Robotic Ablation of Atrial Fibrillation
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
05:43Determination of Continuity Index Values in Atrial Fibrillation Ablation with Proactive Esophageal Cooling
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
09:17High-Resolution Endocardial and Epicardial Optical Mapping in a Sheep Model of Stretch-Induced Atrial Fibrillation